Tibial component shelf life was examined as a contributory factor of the in vivo failure of gamma sterilized prosthetic knee replacements. One hundred eighty-eight Synatomic total knee replacements sterilized by gamma irradiation in air were implanted by one surgeon into 147 patients between May 1985 and December 1994. Of these, 135 knees in 105 patients with a mean followup of 5.8 years (range, 2.1-11.3 years) were included in the study. The mean shelf life of the implants was 3.6 years (range, 0.1-10.7 years). Clinical failure for this study was defined as component retrieval resulting from polyethylene degradation. The knee components were divided into three different groups determined by their shelf storage durations of 0 to 4 years (Group 1, 93 components), between 4 and 8 years (Group 2, 21 components), or greater than 8 to 11 years (Group 3, 21 components). Six prostheses were revised because of polyethylene degradation after a mean implantation time of 2.5 years (range, 1.1-3.8 years). The mean shelf life of these six prostheses was 8.4 years (range, 5.8-9.6 years). Five years after implantation, prostheses that had shelf lives of less than 4 years had a 100% survival rate. Those that had shelf lives of 4 to 8 years before implantation had an 88.6% survival rate, and those prostheses that had shelf lives greater than 8 to 11 years had a 79.2% survival rate.

摘要

胫骨部件的货架期被视为γ射线灭菌的人工膝关节置换术体内失败的一个促成因素。1985年5月至1994年12月期间,一名外科医生将188例在空气中经γ射线辐照灭菌的Synatomic全膝关节置换物植入147例患者体内。其中,对105例患者的135个膝关节进行了研究,平均随访5.8年(范围2.1 - 11.3年)。植入物的平均货架期为3.6年(范围0.1 - 10.7年)。本研究中的临床失败定义为因聚乙烯降解导致的部件取出。膝关节部件根据其货架储存时间分为三个不同组:0至4年(第1组,93个部件)、4至8年(第2组,21个部件)或大于8至11年(第3组,21个部件)。6个假体在平均植入时间2.5年(范围1.1 - 3.8年)后因聚乙烯降解而翻修。这6个假体的平均货架期为8.4年(范围5.8 - 9.6年)。植入后5年,货架期小于4年的假体生存率为100%。植入前货架期为4至8年的假体生存率为88.6%,而货架期大于8至11年的假体生存率为79.2%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验