Burnett R Stephen J, Biggerstaff Scott, Currier Barbara H, Collier John P, Barrack Robert L
Department of Orthopaedic Surgery, Washington University School of Medicine, and Barnes-Jewish Hospital, St Louis, Missouri, USA.
J Arthroplasty. 2007 Aug;22(5):753-8. doi: 10.1016/j.arth.2006.08.007.
This is a report of a unique case of bilateral simultaneous total knee arthroplasties in which one tibial liner failed dramatically, whereas the other liner showed minimal evidence of wear. This unique case allows isolation of component factors as the primary contributing etiology to failure. The differentiating characteristic was the method of sterilization and the shelf life of the polyethylene liner. The insert that showed minimal wear was sterilized with gamma radiation in a barrier package and had a shelf life of less than 1 year, whereas the insert that failed dramatically was sterilized in air and had a shelf life of more than 5 years. This case provides a dramatic example of the potential detrimental effects of manufacturing details on the performance of orthopedic implants, particularly polyethylene inserts.
这是一篇关于双侧同时进行全膝关节置换术的独特病例报告。在该病例中,一侧胫骨衬垫严重失效,而另一侧衬垫磨损迹象极小。这一独特病例能够将部件因素分离出来,作为导致失败的主要病因。其区别特征在于聚乙烯衬垫的灭菌方法和保质期。磨损极小的衬垫采用伽马射线在阻隔包装中灭菌,保质期不到1年;而严重失效的衬垫在空气中灭菌,保质期超过5年。该病例有力地证明了制造细节对骨科植入物,尤其是聚乙烯衬垫性能可能产生的有害影响。