Surgical management of the patient with an implanted cardiac device: implications of electromagnetic interference.

作者信息

Madigan J D, Choudhri A F, Chen J, Spotnitz H M, Oz M C, Edwards N

机构信息

Department of Surgery, Columbia University, New York City, New York, USA.

出版信息

Ann Surg. 1999 Nov;230(5):639-47. doi: 10.1097/00000658-199911000-00005.

Abstract

OBJECTIVE

To identify the sources of electromagnetic interference (EMI) that may alter the performance of implanted cardiac devices and develop strategies to minimize their effects on patient hemodynamic status.

SUMMARY BACKGROUND DATA

Since the development of the sensing demand pacemaker, EMI in the clinical setting has concerned physicians treating patients with such devices. Implanted cardiovertor defibrillators (ICDs) and ventricular assist devices (VADs) can also be affected by EMI.

METHODS

All known sources of interference to pacemakers, ICDs, and VADs were evaluated and preventative strategies were devised.

RESULTS

All devices should be thoroughly evaluated before and after surgery to make sure that its function has not been permanently damaged or changed. If electrocautery is to be used, pacemakers should be placed in a triggered or asynchronous mode; ICDs should have arrhythmia detection suspended before surgery. If defibrillation is to be used, the current flow between the paddles should be kept as far away from and perpendicular to the lead system as possible. Both pacemakers and ICDs should be properly shielded if magnetic resonance imaging, positron emission tomography, or radiation therapy is to be used. The effect of EMI on VADs depends on the model. Magnetic resonance imaging adversely affects all VADs except the Abiomed VAD, and therefore its use should be avoided in this population of patients.

CONCLUSIONS

The patient with an implanted cardiac device can safely undergo surgery as long as certain precautions are taken.

摘要

相似文献

2
Effects of electromagnetic interference on implanted cardiac devices and their management.
Cardiol Rev. 2007 Nov-Dec;15(6):304-9. doi: 10.1097/CRD.0b013e31813e0ba9.
3
Implications of pacemakers and implantable cardioverter defibrillators in urological practice.
J Urol. 2011 Oct;186(4):1198-205. doi: 10.1016/j.juro.2011.02.2697. Epub 2011 Aug 16.
4
Electromagnetic interference in cardiac rhythm management devices.
AACN Clin Issues. 2004 Jul-Sep;15(3):391-403. doi: 10.1097/00044067-200407000-00007.
5
7
Implantable rhythm devices and electromagnetic interference: myth or reality?
Expert Rev Cardiovasc Ther. 2008 Jul;6(6):823-32. doi: 10.1586/14779072.6.6.823.
8
[Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].
Herzschrittmacherther Elektrophysiol. 2022 Sep;33(3):297-304. doi: 10.1007/s00399-022-00875-7. Epub 2022 Jul 4.
10
Electromagnetic interference and implanted cardiac devices: the nonmedical environment (part I).
Clin Cardiol. 2012 May;35(5):276-80. doi: 10.1002/clc.21998. Epub 2012 Apr 26.

引用本文的文献

1
Safe and effective intraosseous basivertebral nerve radiofrequency neurotomy in a patient with a permanent pacemaker.
Interv Pain Med. 2025 Mar 20;4(1):100565. doi: 10.1016/j.inpm.2025.100565. eCollection 2025 Mar.
2
Cardiac interventions on total knee arthroplasty: a national inpatient sample-based analysis.
Arch Orthop Trauma Surg. 2024 May;144(5):2229-2238. doi: 10.1007/s00402-024-05225-5. Epub 2024 Feb 29.
3
Mohs micrographic surgery with magnet suspension of implantable cardioverter defibrillators.
Skin Health Dis. 2024 Jan 6;4(1):e331. doi: 10.1002/ski2.331. eCollection 2024 Feb.
4
Anesthetic Management of Patients with Dilated Cardiomyopathy Undergoing Noncardiac Surgery.
Medicina (Kaunas). 2023 Aug 29;59(9):1567. doi: 10.3390/medicina59091567.
5
Surgical Implications of Implanted Cardiac Rhythm Devices.
J Innov Card Rhythm Manag. 2023 Jan 15;14(1):5309-5312. doi: 10.19102/icrm.2023.14015. eCollection 2023 Jan.
6
Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement.
Cureus. 2022 Sep 21;14(9):e29391. doi: 10.7759/cureus.29391. eCollection 2022 Sep.
7
Pacemaker Failure to Capture Caused by Electrocautery: A Rare Pacemaker Pulse Generator Change Complication.
Cureus. 2022 Aug 22;14(8):e28252. doi: 10.7759/cureus.28252. eCollection 2022 Aug.
8
Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review.
J Interv Card Electrophysiol. 2021 Jun;61(1):1-10. doi: 10.1007/s10840-020-00930-8. Epub 2021 Jan 12.
9
Anesthetic consideration for patients with micra leadless pacemaker.
Ann Card Anaesth. 2020 Oct-Dec;23(4):493-495. doi: 10.4103/aca.ACA_191_19.
10
Electrosurgery and Temperature Increase in Tissue With a Passive Metal Implant.
Front Surg. 2019 Mar 12;6:8. doi: 10.3389/fsurg.2019.00008. eCollection 2019.

本文引用的文献

1
Use of an ultrasonic scalpel as an alternative to electrocautery in patients with pacemakers.
Ann Thorac Surg. 1998 Jun;65(6):1802-4. doi: 10.1016/s0003-4975(98)00105-2.
2
The patient with a pacemaker or related device.
Can J Anaesth. 1996 May;43(5 Pt 2):R24-41. doi: 10.1007/BF03011666.
3
Ultrasonic dissection.
Endosc Surg Allied Technol. 1994 Jun-Aug;2(3-4):181-5.
4
North American Society of Pacing and Electrophysiology policy statement. The NASPE/BPEG defibrillator code.
Pacing Clin Electrophysiol. 1993 Sep;16(9):1776-80. doi: 10.1111/j.1540-8159.1993.tb01809.x.
7
The effects of nuclear magnetic resonance imagers on external and implantable pulse generators.
Pacing Clin Electrophysiol. 1984 Jul;7(4):720-7. doi: 10.1111/j.1540-8159.1984.tb05602.x.
8
Electrocautery and pacemakers: management of the paced patient subject to electrocautery.
Ann Thorac Surg. 1986 Mar;41(3):313-7. doi: 10.1016/s0003-4975(10)62777-4.
9
Effect of magnetic resonance imaging on DDD pacemakers.
Am J Cardiol. 1986 Feb 15;57(6):437-40. doi: 10.1016/0002-9149(86)90768-x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验