The effects of nuclear magnetic resonance imagers on external and implantable pulse generators.

作者信息

Fetter J, Aram G, Holmes D R, Gray J E, Hayes D L

出版信息

Pacing Clin Electrophysiol. 1984 Jul;7(4):720-7. doi: 10.1111/j.1540-8159.1984.tb05602.x.

Abstract

This study evaluates the effect of nuclear magnetic resonance (NMR) scanning on pacemaker function. It must be emphasized that each manufacturer's pulse generators and each pacing modality may behave differently and, therefore, require individual evaluation. According to our results, patients with pacemakers should have their pacing activity monitored continuously during scanning with the NMR 1500 gauss imaging system. External pulse generators should be set to the asynchronous mode and placed outside the NMR image volume but within the radiofrequency (RF) shield. Implanted pacemakers should be verified for type and mode of operation. All implantable pulse generators evaluated reverted from the demand to the asynchronous mode within the magnetic field of the scanner. There was no observable damage to the discrete pacemaker components that were tested. In vivo testing of implantable single-chamber pulse generators did not significantly alter the pacemaker's operating parameters. Changes in stimulation rate analogous to the RF field pulse rate were seen. In single-chamber devices the resultant rate was a multiple of the RF frequency, changing to a value less than the normal asynchronous magnetic rate. With more sophisticated dual-chamber devices the results varied. With VDD pacing during RF scanning, the cardiac stimulation rate increased to a value analogous to the RF field modulation period. More extensive in vivo testing using different models of pulse generators of various manufacturers is needed to identify specific device susceptibility to the RF, time variance and steady-state magnet fields. From these data a comprehensive statement about NMR scanning of patients with implanted pacemakers can be made.

摘要

本研究评估了核磁共振(NMR)扫描对起搏器功能的影响。必须强调的是,每个制造商的脉冲发生器以及每种起搏模式的表现可能有所不同,因此需要进行单独评估。根据我们的结果,植入起搏器的患者在使用NMR 1500高斯成像系统进行扫描期间,其起搏活动应持续受到监测。外部脉冲发生器应设置为异步模式,并放置在NMR图像容积之外但在射频(RF)屏蔽范围内。应核实植入起搏器的类型和工作模式。所评估的所有植入式脉冲发生器在扫描仪磁场内均从按需模式转换为异步模式。对所测试的分立起搏器组件未观察到明显损坏。对植入式单腔脉冲发生器进行的体内测试并未显著改变起搏器的工作参数。观察到刺激率的变化类似于RF场脉冲率。在单腔装置中,最终速率是RF频率的倍数,变为小于正常异步磁速率的值。对于更复杂的双腔装置,结果有所不同。在RF扫描期间进行VDD起搏时,心脏刺激率增加到类似于RF场调制周期的值。需要使用不同制造商的不同型号脉冲发生器进行更广泛的体内测试,以确定特定设备对RF、时间变化和稳态磁场的敏感性。根据这些数据,可以对植入起搏器患者的NMR扫描做出全面的说明。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验