Performance evaluation of a multi-slice CT system.

作者信息

McCollough C H, Zink F E

机构信息

Department of Diagnostic Radiology, Mayo Clinic, Rochester, Minnesota 55905, USA.

出版信息

Med Phys. 1999 Nov;26(11):2223-30. doi: 10.1118/1.598777.

Abstract

Our purpose in this study was to characterize the performance of a recently introduced multi-slice CT scanner (LightSpeed QX/i, Version 1.0, General Electric Medical Systems) in comparison to a single-slice scanner from the same manufacturer (HiSpeed CT/i, Version 4.0). To facilitate this comparison, a refined definition of pitch is introduced which accommodates multi-slice CT systems, yet maintains the existing relationships between pitch, patient dose, and image quality. The following performance parameters were assessed: radiation and slice sensitivity profiles, low-contrast and limiting spatial resolution, image uniformity and noise, CT number and geometric accuracy, and dose. The multi-slice system was tested in axial (1, 2, or 4 images per gantry rotation) and HQ (Pitch = 0.75) and HS (Pitch = 1.5) helical modes. Axial and helical acquisition speed and limiting spatial resolution (0.8-s exposure) were improved on the multi-slice system. Slice sensitivity profiles, image noise, CT number accuracy and uniformity, and low-contrast resolution were similar. In some HS-helical modes, helical artifacts and geometric distortion were more pronounced with a different appearance. Radiation slice profiles and doses were larger on the multi-slice system at all scan widths. For a typical abdomen and pelvis exam, the central and surface body doses for 5-mm helical scans were higher on the multi-slice system by approximately 50%. The increase in surface CTDI values (with respect to the single-slice system) was greatest for the 4 x 1.25-mm detector configuration (190% for head, 240% for body) and least for the 4 x 5-mm configuration (53% for head, 76% for body). Preliminary testing of version 1.1 software demonstrated reduced doses on the multi-slice scanner, where the increase in body surface CTDI values (with respect to the single-slice system) was 105% for the 4 x 1.25-mm detector configuration and 10% for the 4 x 5-mm configuration. In summary, the axial and HQ-helical modes of the multi-slice system provided excellent image quality and a substantial reduction in exam time and tube loading, although at varying degrees of increased dose relative to the single-slice scanner.

摘要

我们开展本研究的目的是,将最近推出的多层CT扫描仪(LightSpeed QX/i,版本1.0,通用电气医疗系统公司)与同一制造商生产的单层扫描仪(HiSpeed CT/i,版本4.0)的性能进行对比。为便于这种对比,引入了螺距的精确界定,该界定适用于多层CT系统,同时保持螺距、患者剂量和图像质量之间的现有关系。评估了以下性能参数:辐射和层厚灵敏度曲线、低对比度和极限空间分辨率、图像均匀性和噪声、CT值和几何精度以及剂量。多层系统在轴向(每机架旋转1、2或4幅图像)以及HQ(螺距=0.75)和HS(螺距=1.5)螺旋模式下进行了测试。多层系统的轴向和螺旋采集速度以及极限空间分辨率(0.8秒曝光)有所提高。层厚灵敏度曲线、图像噪声、CT值准确性和均匀性以及低对比度分辨率相似。在某些HS螺旋模式下,螺旋伪影和几何畸变更为明显,且外观有所不同。在所有扫描宽度下,多层系统的辐射层厚曲线和剂量都更大。对于典型的腹部和骨盆检查,多层系统5毫米螺旋扫描的中心和体表剂量比单层系统高约50%。表面CTDI值(相对于单层系统)的增加在4×1.25毫米探测器配置中最大(头部为190%,身体为240%)而在4×5毫米配置中最小(头部为53%,身体为76%)。对1.1版本软件的初步测试表明,多层扫描仪的剂量有所降低,其中对于4×1.25毫米探测器配置,体表CTDI值的增加(相对于单层系统)为105%,对于4×5毫米配置为10%。总之,多层系统的轴向和HQ螺旋模式提供了出色的图像质量,并大幅缩短了检查时间和降低了球管负荷,尽管相对于单层扫描仪,剂量增加的程度有所不同。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验