Wess G, Reusch C
Clinic for Small Animal Internal Medicine, University of Zurich, Switzerland.
J Am Vet Med Assoc. 2000 Jan 15;216(2):203-9. doi: 10.2460/javma.2000.216.203.
To evaluate clinical and analytical accuracy of 5 portable blood glucose meters (PBGM) used to measure blood glucose concentrations in dogs and to determine potential sources of error.
Prospective study.
221 dogs.
Venous blood samples were obtained, and results of the 5 PBGM were compared with results of a hexokinase reference method. Agreement among methods was determined by use of error grid analysis and statistical methods.
Accuracy of the PBGM varied with glucose concentration of the sample. The largest differences between results of the PBGM and results of the reference method were obtained with samples with high glucose concentrations; 4 PBGM tended to underestimate and 1 PBGM tended to overestimate the true glucose concentration. Absolute differences between results of the PBGM and results of the reference method were small for samples with low glucose concentrations and samples with concentrations in the reference range. None of the PBGM yielded measurements that would result in clinically unacceptable errors. Within-run and between-day precision was good for all PBGM, and results were not affected by use of EDTA or heparin to anticoagulate blood. Readings of the PBGM were significantly higher for blood samples with low Hct than for samples with normal Hct. For 3 PBGM, samples < 3 microliters resulted in inaccurate measurements.
Results suggest that currently available PBGM are sufficiently accurate for use in clinical practice to determine blood glucose concentrations in dogs.
评估5种用于测量犬血糖浓度的便携式血糖仪(PBGM)的临床和分析准确性,并确定潜在的误差来源。
前瞻性研究。
221只犬。
采集静脉血样本,将5种PBGM的检测结果与己糖激酶参考方法的结果进行比较。采用误差网格分析和统计方法确定各方法之间的一致性。
PBGM的准确性随样本葡萄糖浓度而异。PBGM结果与参考方法结果之间的最大差异出现在高葡萄糖浓度样本中;4种PBGM倾向于低估,1种PBGM倾向于高估真实葡萄糖浓度。对于低葡萄糖浓度样本和参考范围内浓度的样本,PBGM结果与参考方法结果之间的绝对差异较小。没有一种PBGM的测量结果会导致临床上不可接受的误差。所有PBGM的批内和日间精密度良好,结果不受使用乙二胺四乙酸(EDTA)或肝素抗凝血液的影响。血细胞比容(Hct)低的血样PBGM读数显著高于Hct正常的血样。对于3种PBGM,体积小于3微升的样本会导致测量不准确。
结果表明,目前可用的PBGM在临床实践中用于测定犬血糖浓度具有足够的准确性。