Suppr超能文献

用于测量犬血糖浓度的人用便携式血糖仪、兽用便携式血糖仪和自动化学分析仪的比较。

Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs.

作者信息

Johnson Beth M, Fry Michael M, Flatland Bente, Kirk Claudia A

机构信息

Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Tennessee, Knoxville, TN 37996, USA.

出版信息

J Am Vet Med Assoc. 2009 Dec 1;235(11):1309-13. doi: 10.2460/javma.235.11.1309.

Abstract

OBJECTIVE

To compare blood glucose concentrations measured with 2 portable blood glucose meters (PBGMs) validated for use in dogs (PBGM-D) and humans (PBGM-H) and an automated chemistry analyzer.

DESIGN

Validation study.

SAMPLE POPULATION

92 samples of fresh whole blood and plasma from 83 dogs with various diseases.

PROCEDURES

Each PBGM was used to measure whole blood glucose concentration, and the automated analyzer was used to measure plasma glucose concentration. Passing-Bablok linear regression and Bland-Altman plots were used to determine correlations and bias between the PBGMs and the automated analyzer. Calculated acceptability limits based on combined inherent instrument imprecision were used with Bland-Altman plots to determine agreement. Clinical relevance was assessed via error grid analysis.

RESULTS

Although correlation between results of both PBGMs and the standard analyzer was > 0.90, disagreement was greater than could be explained by instrument imprecision alone. Mean difference between PBGM-H and chemistry-analyzer values was -15.8 mg/dL. Mean difference between PBGM-D and chemistry-analyzer values was 2.4 mg/dL. Linear regression analysis revealed proportional bias of PBGM-H (greater disagreement at higher glucose concentrations); no proportional bias was detected for PBGM-D. No constant bias was detected for either PBGM. Error grid analysis revealed all measurements from both PBGMs were within zones without an anticipated effect on clinical outcome.

CONCLUSIONS AND CLINICAL RELEVANCE

Neither PBGM had exact agreement with the automated analyzer; however, the disagreement detected did not have serious clinical consequences. Our findings stressed the importance of using the same device for monitoring trends in dogs and using instrument-specific reference ranges.

摘要

目的

比较两种经验证可用于犬类(PBGM-D)和人类(PBGM-H)的便携式血糖仪以及一台自动化学分析仪所测得的血糖浓度。

设计

验证研究。

样本群体

来自83只患有各种疾病的犬的92份新鲜全血和血浆样本。

程序

使用每种便携式血糖仪测量全血葡萄糖浓度,使用自动分析仪测量血浆葡萄糖浓度。采用Passing-Bablok线性回归和Bland-Altman图来确定便携式血糖仪与自动分析仪之间的相关性和偏差。将基于仪器固有不精密度合并计算得出的可接受限度用于Bland-Altman图以确定一致性。通过误差网格分析评估临床相关性。

结果

尽管两种便携式血糖仪与标准分析仪的结果之间的相关性均>0.90,但差异大于仅由仪器不精密度所能解释的范围。PBGM-H与化学分析仪值之间的平均差异为-15.8mg/dL。PBGM-D与化学分析仪值之间的平均差异为2.4mg/dL。线性回归分析显示PBGM-H存在比例偏差(在较高葡萄糖浓度下差异更大);未检测到PBGM-D存在比例偏差。两种便携式血糖仪均未检测到恒定偏差。误差网格分析显示,两种便携式血糖仪的所有测量值均在对临床结果无预期影响的区域内。

结论与临床相关性

两种便携式血糖仪与自动分析仪均未完全一致;然而,所检测到的差异并无严重临床后果。我们的研究结果强调了在监测犬类血糖变化趋势时使用同一设备以及使用特定仪器参考范围的重要性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验