Med Biol Eng Comput. 1999 Jul;37(4):517-22. doi: 10.1007/BF02513339.
Episodes of serious but infrequently occurring cardiac arrhythmias can be difficult to detect and analyse, even with modern Holter monitoring. A previous diagnostic pacemaker developed by this group provided VVI pacing therapy and recorded intracardiac ECG signals but had no atrial sensing or impedance measuring capability. A new external diagnostic pacemaker system is described that has been developed to assist in diagnosing intermittent arrhythmias by selectively recording intracardiac signals. Unlike other ambulatory monitors, in addition to recording ECG, the device combines VDD pacing therapy with the capability of monitoring and recording intracardiac impedance and pressure waveforms through a temporary intracardiac lead. A PCMCIA memory card allows storage of 48 arrhythmic events of 21 seconds each. Twelve seconds of waveform are retained before the event occurs and nine seconds after. Data retrieval and processing is performed with a PC which reconstructs each waveform for display. The ECG provides data on cardiac rhythm while cardiac function is inferred from the haemodynamic signals. During simulated trials, 14 event types were presented to the system. All events were successfully detected and recorded. During in vivo clinical tests 83 waveform recordings were made. Impedance fluctuations with typical peak-to-peak values of 64 ohms were successfully recorded.
即使使用现代动态心电图监测,严重但不常发生的心律失常发作也可能难以检测和分析。该团队之前开发的一款诊断性起搏器提供VVI起搏治疗并记录心内心电图信号,但没有心房感知或阻抗测量功能。本文介绍了一种新的体外诊断起搏器系统,该系统通过选择性记录心内信号来辅助诊断间歇性心律失常。与其他动态监测仪不同,除了记录心电图外,该设备还将VDD起搏治疗与通过临时心内导联监测和记录心内阻抗及压力波形的功能相结合。一张PCMCIA存储卡可存储48个心律失常事件,每个事件时长21秒。事件发生前保留12秒波形,事件发生后保留9秒波形。使用一台计算机进行数据检索和处理,该计算机重建每个波形以供显示。心电图提供心律数据,而心脏功能则从血流动力学信号中推断得出。在模拟试验中,向该系统呈现了14种事件类型。所有事件均被成功检测和记录。在体内临床试验中,进行了83次波形记录。成功记录了典型峰峰值为64欧姆的阻抗波动。