Feasibility of measuring relative right ventricular volumes and ejection fraction with implantable rhythm control devices.

作者信息

Chirife R, Ortega D F, Salazar A I

机构信息

Institute of Cardiology, National Academy of Medicine, Buenos Aires, Argentina.

出版信息

Pacing Clin Electrophysiol. 1993 Aug;16(8):1673-83. doi: 10.1111/j.1540-8159.1993.tb01038.x.

Abstract

Ejection fraction (EF), the ratio between stroke volume (SV) and end-diastolic volume (EDV), is a valuable contractility indicator. Unlike SV, the Frank-Starling effect is automatically compensated in the calculation of EF. It was the aim of this study to evaluate the physiological behavior of impedance derived measurements of relative right ventricular (RV) volumes and EF, obtained with standard pacing leads. Seven patients were evaluated at the time of pacemaker implant or replacement. Since no absolute standard of comparison was available for RV volumes, the value of the measurements was assessed by observing their behavior under cardio-circulatory challenges. A 2.5-kHz carrier was fed to the ring and tip electrodes of standard bipolar pacing leads and the resulting voltage was digitized and stored. The peak-to-peak voltage (PPV) of the carrier at the time of QRS was used as EDV, and the largest PPV as end-systolic volume (ESV). Relative SV was the difference between EDV and ESV, and EF = SV/EDV x 100. Pacing was used to reduce EDV, and the effect of contractility was tested with isometric hand grip, recumbent leg exercise, or isoproterenol drip. Only minimal changes in EF were noted during incremental pacing; relative SV and EDV decreased as expected; and EF increased significantly during contractility challenges. A high correlation coefficient was observed between EDV and SV changes induced by incremental pacing at rest (r values from 0.62 to 0.98, P from < 0.01 to 0.001). The study revealed that impedance volumetry, utilizing conventional bipolar pacing leads, yields useful hemodynamic data related to EDV, ESV, and EF. Given the simplicity of the method, it is reasonable to conclude on the feasibility of using said impedance derived hemodynamic parameters in implantable rhythm control devices.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验