Pressure measurements during injection of corticosteroids: in vivo studies.

作者信息

Paul S, Hoey M F, Egbert J E

机构信息

Department of Biomedical Engineering, University of Minnesota, Minneapolis 55455, USA.

出版信息

Med Biol Eng Comput. 1999 Sep;37(5):645-51. doi: 10.1007/BF02513361.

Abstract

Intralesional injection of corticosteroids is an effective treatment for capillary hemangiomas. Complications include embolisation of corticosteroid particles into the ocular circulation resulting in permanent loss of vision. This research is aimed at developing an injection cannula and monitoring system to prevent such inadvertent embolisation. A cannula has been designed to simultaneously estimate the pressure at its tip and the flow rate during injection. The estimation technique has previously been validated using an in vitro model. In this study, the cannula was tested in vivo with canine liver at injection flow rates of 2.5 to 21 ml min-1. The pressure generated in the tissue during injection was calculated using the technique developed. This was compared with direct in situ pressure measurements made with a coaxial outer cannula. The mean calculated pressure was seen to be linearly related to the mean measured pressure with a slope of 0.97, correlation coefficient of 0.99 and standard error of 2.74 mmHg. Similar trends were observed between the maximum calculated and maximum measured injection pressure: slope = 1.0, r = 0.99 and standard error = 5.54 mmHg. The estimation of the mean and maximum pressure from the cannula and monitoring system was accurate in canine liver. High pressures close to 250 mmHg were generated in tissues during injection.

摘要

瘤内注射皮质类固醇是治疗毛细血管瘤的一种有效方法。并发症包括皮质类固醇颗粒栓塞进入眼循环,导致永久性视力丧失。本研究旨在开发一种注射套管和监测系统,以防止此类意外栓塞。已设计出一种套管,可在注射过程中同时估计其尖端的压力和流速。该估计技术先前已通过体外模型进行验证。在本研究中,该套管在犬肝脏上进行了体内测试,注射流速为2.5至21毫升/分钟。使用所开发的技术计算注射过程中组织内产生的压力。将其与使用同轴外套管进行的直接原位压力测量结果进行比较。计算得到的平均压力与测量得到的平均压力呈线性相关,斜率为0.97,相关系数为0.99,标准误差为2.74毫米汞柱。在计算得到的最大注射压力和测量得到的最大注射压力之间也观察到类似趋势:斜率 = 1.0,r = 0.99,标准误差 = 5.54毫米汞柱。该套管和监测系统对犬肝脏平均压力和最大压力的估计是准确的。注射过程中组织内产生了接近250毫米汞柱的高压。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验