Cahill M D, Humphrey V F
Department of Physics, University of Bath, Bath, United Kingdom.
Ultrasound Med Biol. 2000 Mar;26(3):433-40. doi: 10.1016/s0301-5629(99)00153-2.
Safety parameters for diagnostic ultrasound scanners are calculated from measurements made in water, which are derated to account for the attenuation of tissues. Sound is attenuated less by water than by tissue, and so the effects of nonlinear propagation are greater in water. This study compares mechanical index (MI) and derated intensity with the analogous quantities in idealised soft tissue, for simplified models of scanners with source amplitudes up to 2.5 MPa. As expected, MI is much smaller than implied by linear extrapolation from low-amplitude measurements but, in a system with moderate gain, the reduction in tissue is commensurate with that in water, MI and derated intensity underestimating the values in tissue by at most 20%. Determining MI at the location of peak negative pressure halves the error. In high gain systems, however, MI can be less than 60% of the value in tissue.
诊断超声扫描仪的安全参数是根据在水中的测量值计算得出的,这些测量值会进行降额处理以考虑组织的衰减。声音在水中的衰减比在组织中少,因此非线性传播在水中的影响更大。本研究针对源振幅高达2.5MPa的简化扫描仪模型,将机械指数(MI)和降额强度与理想化软组织中的类似量进行了比较。正如预期的那样,MI远小于从低振幅测量值进行线性外推所暗示的值,但在增益适中的系统中,组织中的降低与水中的降低相当,MI和降额强度低估组织中的值最多20%。在峰值负压位置确定MI可将误差减半。然而,在高增益系统中,MI可能小于组织中值的60%。