通过干涉测量法测量角膜厚度。

Corneal thickness measured by interferometry.

作者信息

Green D G, Frueh B R, Shapiro J M

出版信息

J Opt Soc Am. 1975 Feb;65(2):119-23. doi: 10.1364/josa.65.000119.

Abstract

An optical method for measuring the thickness of transparent structures has been developed, and has been used to measure, in vivo, the thickness of the human cornea. The thickness is measured by placing the anterior surface of the cornea at the focus of a beam of coherent laser light and then measuring the spacing between the interference fringes generated by the reflected light. The thickness is then calculated from the fringe spacing. The method has been used to measure corneal thickness in frog and human corneas. These measurements have been correlated with histologic and pachometer measurements of corneal thickness. A significant capability of this technique is to measure the thickness of optically opaque corneas.

摘要

一种用于测量透明结构厚度的光学方法已被开发出来,并已用于在活体中测量人角膜的厚度。通过将角膜前表面置于一束相干激光束的焦点处,然后测量反射光产生的干涉条纹之间的间距来测量厚度。然后根据条纹间距计算厚度。该方法已用于测量青蛙和人角膜的厚度。这些测量结果已与角膜厚度的组织学测量和角膜厚度计测量结果相关联。这项技术的一个重要能力是测量光学不透明角膜的厚度。