Suppr超能文献

用于分枝杆菌培养的BACTEC MGIT 960系统评估

Evaluation of the BACTEC MGIT 960 system for the recovery of mycobacteria.

作者信息

Kanchana M V, Cheke D, Natyshak I, Connor B, Warner A, Martin T

机构信息

Division of Clinical Microbiology, Royal University Hospital, Saskatoon, Saskatchewan, Canada.

出版信息

Diagn Microbiol Infect Dis. 2000 May;37(1):31-6. doi: 10.1016/s0732-8893(99)00151-0.

Abstract

We evaluated the BACTEC MGIT 960 system, which is a fully automated, non-invasive, continuous monitoring system for the growth and detection of mycobacteria. Including respiratory and other specimens, 1,742 specimens were processed and inoculated into the BACTEC MGIT 960 and the BACTEC 460 TB Systems, as well as onto Lowenstein-Jensen (L-J) media. A total of 104 isolates of mycobacteria were recovered from all culture systems. This included Mycobacterium tuberculosis complex, Mycobacterium avium-intracellulare (MAI) complex and other mycobacteria (MOTT). The isolation rates for M. tuberculosis complex and MAI complex were comparable for the BACTEC 460 (54.8% and 13.5%) and the BACTEC MGIT 960 (51.9% and 13.5%). The overall isolation rate was less for BACTEC MGIT 960 (76.9%) which was due to lesser number of MOTT isolates recovered from this system. The mean times to detection (TTD) for all mycobacteria were 9.3 days for the BACTEC MGIT, 14.6 days for the BACTEC 460 and 21.6 days for L-J. A significant difference was observed when TTD was tested in relation to degree of positivity in smears, with the BACTEC MGIT maintaining the short TTD even with less number of bacilli in the smear. The contamination rates were, 6.4% for BACTEC MGIT, 2.9% for BACTEC 460 and 12.1% for L-J medium. The BACTEC MGIT 960 system shows performances comparable to the BACTEC 460 and seems to be a dependable, user friendly system.

摘要

我们评估了BACTEC MGIT 960系统,这是一种用于分枝杆菌生长和检测的全自动、非侵入性连续监测系统。包括呼吸道和其他标本在内,共处理了1742份标本,并接种到BACTEC MGIT 960、BACTEC 460 TB系统以及罗-琴(L-J)培养基上。从所有培养系统中共分离出104株分枝杆菌。这包括结核分枝杆菌复合群、鸟分枝杆菌-胞内分枝杆菌(MAI)复合群和其他分枝杆菌(非结核分枝杆菌)。结核分枝杆菌复合群和MAI复合群在BACTEC 460(54.8%和13.5%)和BACTEC MGIT 960(51.9%和13.5%)中的分离率相当。BACTEC MGIT 960的总体分离率较低(76.9%),这是因为从该系统中回收的非结核分枝杆菌数量较少。BACTEC MGIT对所有分枝杆菌的平均检测时间(TTD)为9.3天,BACTEC 460为14.6天,L-J为21.6天。当检测TTD与涂片阳性程度的关系时,观察到显著差异,即使涂片杆菌数量较少,BACTEC MGIT的TTD仍保持较短。污染率方面,BACTEC MGIT为6.4%,BACTEC 460为2.9%,L-J培养基为12.1%。BACTEC MGIT 960系统表现出与BACTEC 460相当的性能,似乎是一个可靠且用户友好的系统。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验