Silver M D
Bio-Imaging Research, Inc., Lincolnshire, Illinois 60069, USA.
Med Phys. 2000 Apr;27(4):773-4. doi: 10.1118/1.598939.
A method for including redundant data in fan-beam computed tomography (CT) is presented. It is a natural extension of the Parker [Med. Phys. 9, 254-257 (1982)] short-scan approach applied to divergent fan-beam (or cone-beam) data when the data set covers between the minimally complete set of 180 degrees plus fan angle and 360 degrees. A virtual fan angle is introduced whose value is the difference between the angular range of the data collected and 180 degrees. Parker-weights are then applied as if the field-of-view is spanned by the virtual fan angle.
本文提出了一种在扇束计算机断层扫描(CT)中纳入冗余数据的方法。它是帕克[《医学物理》9, 254 - 257 (1982)]短扫描方法的自然扩展,当数据集覆盖范围在最小完整的180度加扇角到360度之间时,该方法应用于发散扇束(或锥束)数据。引入了一个虚拟扇角,其值为收集到的数据的角度范围与180度之间的差值。然后应用帕克权重,就好像视野是由虚拟扇角覆盖的一样。