Jin W, Dong Q, Yu D, Ye X
Laboratory of Analytical Science, School of Chemistry, Shandong University, Jinan, China.
Electrophoresis. 2000 May;21(8):1535-9. doi: 10.1002/(SICI)1522-2683(20000501)21:8<1535::AID-ELPS1535>3.0.CO;2-8.
Capillary zone electrophoresis was employed for the determination of myoglobin in human urine using end-column amperometric detection with a carbon fiber microelectrode at a constant potential of 1.80 V vs. saturated calomel electrode (SCF). The optimum conditions of separation and detection are: 3.73 x 10-4 mol/L sodium diethyl malonyl urea (barbitone sodium), 1.34 x 10-4 mol/L HCl for the buffer solution, 20 kV for separation voltage, 5 kV and 5 s for injection voltage and injection time, respectively. The limit of detection is 4.4 x 10-8 mol/L or 84 amole signal to noise (S/N = 2). The relative standard deviation is 2.9% for the migration time and 2.5% for the electrophoretic peak current. The method can be used for the determination of myoglobin in human urine. The samples can be directly injected and need no pretreatment. The method is also rapid, less than 2 min, and has a recovery rate of 94-106%.
采用毛细管区带电泳法,以碳纤维微电极在相对于饱和甘汞电极(SCF)为1.80 V的恒定电位下进行柱端安培检测,测定人尿中的肌红蛋白。分离和检测的最佳条件为:缓冲溶液为3.73×10⁻⁴ mol/L丙二酰脲二乙钠(巴比妥钠)、1.34×10⁻⁴ mol/L HCl,分离电压为20 kV,进样电压和进样时间分别为5 kV和5 s。检测限为4.4×10⁻⁸ mol/L或84 amole信噪比(S/N = 2)。迁移时间的相对标准偏差为2.9%,电泳峰电流的相对标准偏差为2.5%。该方法可用于测定人尿中的肌红蛋白。样品可直接进样,无需预处理。该方法也很快,不到2分钟,回收率为94 - 106%。