Krieglstein G K, Waller W K, Reimers H, Langham M E
Albrecht Von Graefes Arch Klin Exp Ophthalmol. 1976 May 26;199(3):223-9. doi: 10.1007/BF00417291.
The pneumotonograph was evaluated for its capability to record and measure intraocular pressure through soft contact lenses. In a series of patients with chronic corneal diseases treated with soft contact lenses, good pressure and intraocular pulsations were recorded both with and without the soft lenses. Areas of scarred cornea give incorrect values, whereas areas of corneal edema give values similar to those obtained in the presence of a soft lens and are believed to reflect accurate and reliable pressure readings. In normal eyes, it has been found that the presence of a plain soft lens does not modify the accuracy of pressure measurement. High values do occur, however, when the refraction of the lens exceeds eight diopters.
对气动眼压计通过软性隐形眼镜记录和测量眼压的能力进行了评估。在一系列使用软性隐形眼镜治疗的慢性角膜疾病患者中,无论有无软性隐形眼镜,均记录到了良好的眼压和眼内搏动。角膜瘢痕区域给出的数值不准确,而角膜水肿区域给出的数值与使用软性隐形眼镜时获得的数值相似,并且被认为反映了准确可靠的眼压读数。在正常眼睛中,已发现普通软性隐形眼镜的存在不会改变眼压测量的准确性。然而,当镜片的屈光度超过8时,确实会出现高值。