Zeri Fabrizio, De Cusatis Mario, Lupelli Luigi, Swann Peter Graham
Degree Course in Optics and Optometry, Department of Sciences - Roma TRE University, Rome, Italy; Vision Sciences Department, Istituto Benigno Zaccagnini, Bologne, Italy; School of Life and Health Sciences, Aston University, Birmingham, UK.
Degree Course in Optics and Optometry, Department of Sciences - Roma TRE University, Rome, Italy.
J Optom. 2016 Jul-Sep;9(3):182-8. doi: 10.1016/j.optom.2015.09.001. Epub 2015 Oct 21.
To investigate if the accuracy of intraocular pressure (IOP) measurements using rebound tonometry over disposable hydrogel (etafilcon A) contact lenses (CL) is affected by the positive power of the CLs.
The experimental group comprised 26 subjects, (8 male, 18 female). IOP measurements were undertaken on the subjects' right eyes in random order using a Rebound Tonometer (ICare). The CLs had powers of +2.00D and +6.00D. Measurements were taken over each contact lens and also before and after the CLs had been worn.
The IOP measure obtained with both CLs was significantly lower compared to the value without CLs (t test; p<0.001) but no significant difference was found between the two powers of CLs.
Rebound tonometry over positive hydrogel CLs leads to a certain degree of IOP underestimation. This result did not change for the two positive lenses used in the experiment, despite their large difference in power and therefore in lens thickness. Optometrists should bear this in mind when measuring IOP with the rebound tonometer over plus power contact lenses.