Moissonnier P, Bordeau W, Delisle F, Devauchelle P
Ecole Nationale Vétérinaire d'Alfort, 7 Avenue du Gle De Gaulle, Maisons Alfort, 94700, France.
Res Vet Sci. 2000 Jun;68(3):243-7. doi: 10.1053/rvsc.1999.0370.
The accuracy testing of a new CT -guided stereotactic device was undertaken via an experimental study. Ten dog cadaver heads were submitted to a CT examination and a brain target was determined (coordinates: x1, y1, z1). 0.3 ml of a radiopaque paste was injected in this chosen point. The head was analysed twice by CT and the centre of the injection (x2, y2, z2) was measured on slices. Tridimensional statistical analysis of the variance of the two points was carried out. The procedure always led to the injection of the radiopaque paste and created an intracerebral mass which gave a measurable image on CT examination. The distance between the original target and the centre of the injection was 2.9 +/- 1.08 mm. It is concluded that this new device allows the opportunity to perform stereotactic CT -guided brain biopsy for all unknown brain lesions exceeding 6 mm in diameter in the dog.
通过一项实验研究对一种新型CT引导立体定向装置进行了准确性测试。对10个犬类尸体头部进行CT检查并确定脑靶点(坐标:x1、y1、z1)。在该选定部位注射0.3毫升不透射线的糊剂。头部进行了两次CT分析,并在切片上测量注射中心(x2、y2、z2)。对这两个点的方差进行了三维统计分析。该操作总能成功注射不透射线的糊剂并形成脑内肿块,在CT检查中产生可测量的图像。原始靶点与注射中心之间的距离为2.9±1.08毫米。结论是,这种新装置为对犬类直径超过6毫米的所有不明脑病变进行CT引导下立体定向脑活检提供了机会。