Krause K, Hüdepohl G, Fees H, Rassow J
Strahlentherapie. 1978 Jan;154(1):29-37.
The mode of correction of the isodose curves from telecentric electron pendulous technique using a constant patient radius rp = 30 cm (Isodosenatlas, Siemens, 1973) is represented with regard to its application in patients with diverging surface contours. Correction is possible by two different methods: 1st by experimental determination of an air gap factor for the shift of isodoses, and 2nd by two factors depending on the focus-skin distance and on the angle of incidence of the electron beam. Determination of the factors is performed either by means of fixed fields measured by vertical and at oblique incidence of the beam and a depth dose distribution measured at the central axis, with oblique incidence of the electrons.
利用恒定患者半径rp = 30 cm的远心电子摆动技术(《等剂量图谱》,西门子,1973年)对等剂量曲线进行校正的方法,针对其在表面轮廓发散的患者中的应用进行了阐述。校正可通过两种不同方法实现:第一种是通过实验确定等剂量线位移的气隙因子,第二种是通过两个分别取决于焦皮距和电子束入射角的因子。这些因子的确定可通过以下方式进行:利用电子束垂直入射和斜入射时测量的固定野,以及在电子斜入射时在中心轴处测量的深度剂量分布。