[Contribution to deep electron pendulous therapy. VIII. Communication: concerning the problem of diverging contours in telecentric electron pendulous irradiation using the electron energies 10 MeV and 20 MeV (author's transl)].

作者信息

Krause K, Hüdepohl G, Fees H, Rassow J

出版信息

Strahlentherapie. 1978 Jan;154(1):29-37.

PMID:415385
Abstract

The mode of correction of the isodose curves from telecentric electron pendulous technique using a constant patient radius rp = 30 cm (Isodosenatlas, Siemens, 1973) is represented with regard to its application in patients with diverging surface contours. Correction is possible by two different methods: 1st by experimental determination of an air gap factor for the shift of isodoses, and 2nd by two factors depending on the focus-skin distance and on the angle of incidence of the electron beam. Determination of the factors is performed either by means of fixed fields measured by vertical and at oblique incidence of the beam and a depth dose distribution measured at the central axis, with oblique incidence of the electrons.