Suppr超能文献

Accurate mass liquid chromatography/mass spectrometry on orthogonal acceleration time-of-flight mass analyzers using switching between separate sample and reference sprays. 1. Proof of concept.

作者信息

Eckers C, Wolff J C, Haskins N J, Sage A B, Giles K, Bateman R

机构信息

SmithKline Beecham Pharmaceuticals, New Frontiers Science Park North, Harlow, Essex, UK.

出版信息

Anal Chem. 2000 Aug 15;72(16):3683-8. doi: 10.1021/ac000448i.

Abstract

This paper describes the use of two separate electrosprays for introducing sample and reference for accurate mass liquid chromatography/mass spectrometry (LC/MS) on an orthogonal acceleration time-of-flight mass analyzer. This is carried out using an adaptation of the multiplexed electrospray ion source in which only two of the sprays are utilized. Results are shown for the positive ion detection of trace-level components in complex matrixes and good mass accuracies are obtained, even for very low level components. An example of accurate mass measurements obtained using negative ion LC/MS is also shown. To obtain additional structural information, an example of cone voltage fragmentation is included and shows that good mass accuracy can be obtained for both precursor and fragment ions.