Köfeler Harald C, Gross Michael L
Department of Chemistry, Washington University, St. Louis, Missouri 63130, USA.
J Am Soc Mass Spectrom. 2005 Mar;16(3):406-8. doi: 10.1016/j.jasms.2004.11.022.
The aim of this work is to evaluate quadrupole/time-of-flight (QTOF) mass spectrometry for simultaneous measurement of accurate mass and quantification of a target by using a stable isotopically labeled internal standard. Mixtures of caffeine and (13)C(3)-caffeine (internal standard) at different concentration ratios were analyzed by capillary HPLC/QTOF. A calibration plot for quantification is linear over a factor of 20. Without invoking any correction scheme, the mass accuracy seriously degraded when the ratio of the mass standard to the test compound was not unity. The accuracy could be restored to approximately 2 ppm by using a quadratic function to correct the measured mass as a function of the measured signal ratio of target and internal calibrant.
本研究的目的是评估四极杆/飞行时间(QTOF)质谱仪,通过使用稳定同位素标记的内标物同时测量精确质量和对目标物进行定量。采用毛细管高效液相色谱/四极杆飞行时间质谱联用仪(HPLC/QTOF)分析了不同浓度比的咖啡因和(13)C(3)-咖啡因(内标物)混合物。定量校准曲线在20倍的范围内呈线性。在不采用任何校正方案的情况下,当质量标准物与测试化合物的比例不为1时,质量准确度会严重下降。通过使用二次函数根据目标物与内标物的测量信号比校正测量质量,可将准确度恢复至约2 ppm。