斯旺森(Swanson)和萨特(Sutter)掌指关节植入物的计算机模型分析

Computer model analysis of the Swanson and Sutter metacarpophalangeal joint implants.

作者信息

Williams N W, Penrose J M, Hose D R

机构信息

Department of Plastic Surgery, Northern General Hospital, Sheffield, UK.

出版信息

J Hand Surg Br. 2000 Apr;25(2):212-20. doi: 10.1054/jhsb.1999.0352.

Abstract

A representative model which mimics the behaviour of Silastic finger metacarpophalangeal joint implants was constructed using a finite element software package. The modelled implants were moved through a range of flexion, lateral deviation and a combination of both. Pistoning of both implants stems occurred within the modelled medullary cavities. For equivalent flexion angles, the Sutter implant produced a higher stress field than the Swanson implant, and the field was positioned at the central hinge mechanism. In both implants, lateral deviation increased the internal stress concentrations more than when pure flexion was applied. Overall the Swanson style of implant had lower stress magnitudes than the Sutter implant, and it is predicted that the Sutter implant will be more likely to fail than the Swanson. The failure mode for the Sutter implant would be at the central hinge region. The Swanson implant is likely to fail at the central hinge-stem interface regions.

摘要

使用有限元软件包构建了一个模拟硅橡胶手指掌指关节植入物行为的代表性模型。对模拟植入物进行一系列的屈曲、侧偏以及两者结合的动作。两个植入物柄在模拟的髓腔内发生活塞运动。对于同等的屈曲角度,萨特植入物产生的应力场比斯旺森植入物更高,且该应力场位于中央铰链机构处。在两种植入物中,侧偏比单纯屈曲时更能增加内部应力集中。总体而言,斯旺森式植入物的应力大小低于萨特植入物,预计萨特植入物比斯旺森植入物更易失效。萨特植入物的失效模式将出现在中央铰链区域。斯旺森植入物可能在中央铰链 - 柄界面区域失效。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验