Lawrie G M, Stacey R B, Wright J S
Anaesth Intensive Care. 1975 Nov;3(4):317-20. doi: 10.1177/0310057X7500300406.
Cardiac output has been measured in infants by thermodilution during and after surgery. A method employing commerically available equipment has been developed. An animal model has been devised whereby right heart outputs of 210 ml to 2240 ml, measured by thermodilution, could be compared with left heart outputs measured by an aortic by-pass. The results of this study suggest that thermodilution is a simple, accurate method for measurement of cardiac output in infancy.
在手术期间及术后,已通过热稀释法对婴儿的心输出量进行了测量。现已开发出一种使用市售设备的方法。已设计出一种动物模型,通过该模型,可将通过热稀释法测得的210毫升至2240毫升的右心输出量与通过主动脉旁路测得的左心输出量进行比较。这项研究的结果表明,热稀释法是一种用于测量婴儿心输出量的简单、准确的方法。