Dyson D H, McDonnell W N, Horne J A
Can J Comp Med. 1984 Oct;48(4):425-7.
A model system of feline or small canine cardiac output was used to produce known liquid flow rates in the range of 100 to 750 mL/min for comparison against a thermodilution technique of flow measurement. Thermal indicator size was decided by the thermal time concentration curve detected by the Edwards 9520A cardiac output computer. Ten consecutive readings for each flow were made. Regression analysis and Student's t-test were used to evaluate the results. The computer was found to give good correlation with the accurate flow measured by a graduated cylinder over a period of time (r = 0.99). An error of less than 7% overestimation of flow by thermodilution was found with flows greater than 200 mL/min (p less than 0.05). A significant error of more than 20% overestimation of the actual flow occurred with flows less than 200 mL/min (p less than 0.05). The Edwards 9520A computer was compared to the older Edwards 9510A model by averaged triplicate measurements at six different cardiac outputs in an anesthetized cat. The measurements were not significantly different (p less than 0.01). Thermodilution using an Edwards computer proves to be a promising tool in the measurement of low flows applicable to feline and small canine cardiac outputs.
使用猫或小型犬的心输出量模型系统来产生100至750毫升/分钟范围内已知的液体流速,以便与热稀释法血流测量技术进行比较。热指示剂的大小由爱德华兹9520A心输出量计算机检测到的热时间浓度曲线决定。对每个流速进行连续十次读数。使用回归分析和学生t检验来评估结果。发现该计算机在一段时间内与量筒测量的准确流速具有良好的相关性(r = 0.99)。当流速大于200毫升/分钟时,热稀释法测量的流速高估误差小于7%(p < 0.05)。当流速小于200毫升/分钟时,实际流速高估出现超过20%的显著误差(p < 0.05)。通过在麻醉猫的六种不同心输出量下进行平均三次重复测量,将爱德华兹9520A计算机与旧的爱德华兹9510A型号进行比较。测量结果无显著差异(p < 0.01)。事实证明,使用爱德华兹计算机的热稀释法是测量适用于猫和小型犬心输出量的低流速的一种有前景的工具。