Accuracy of thermodilution measurement of cardiac output in low flows applicable to feline and small canine patients.

作者信息

Dyson D H, McDonnell W N, Horne J A

出版信息

Can J Comp Med. 1984 Oct;48(4):425-7.

Abstract

A model system of feline or small canine cardiac output was used to produce known liquid flow rates in the range of 100 to 750 mL/min for comparison against a thermodilution technique of flow measurement. Thermal indicator size was decided by the thermal time concentration curve detected by the Edwards 9520A cardiac output computer. Ten consecutive readings for each flow were made. Regression analysis and Student's t-test were used to evaluate the results. The computer was found to give good correlation with the accurate flow measured by a graduated cylinder over a period of time (r = 0.99). An error of less than 7% overestimation of flow by thermodilution was found with flows greater than 200 mL/min (p less than 0.05). A significant error of more than 20% overestimation of the actual flow occurred with flows less than 200 mL/min (p less than 0.05). The Edwards 9520A computer was compared to the older Edwards 9510A model by averaged triplicate measurements at six different cardiac outputs in an anesthetized cat. The measurements were not significantly different (p less than 0.01). Thermodilution using an Edwards computer proves to be a promising tool in the measurement of low flows applicable to feline and small canine cardiac outputs.

摘要

使用猫或小型犬的心输出量模型系统来产生100至750毫升/分钟范围内已知的液体流速,以便与热稀释法血流测量技术进行比较。热指示剂的大小由爱德华兹9520A心输出量计算机检测到的热时间浓度曲线决定。对每个流速进行连续十次读数。使用回归分析和学生t检验来评估结果。发现该计算机在一段时间内与量筒测量的准确流速具有良好的相关性(r = 0.99)。当流速大于200毫升/分钟时,热稀释法测量的流速高估误差小于7%(p < 0.05)。当流速小于200毫升/分钟时,实际流速高估出现超过20%的显著误差(p < 0.05)。通过在麻醉猫的六种不同心输出量下进行平均三次重复测量,将爱德华兹9520A计算机与旧的爱德华兹9510A型号进行比较。测量结果无显著差异(p < 0.01)。事实证明,使用爱德华兹计算机的热稀释法是测量适用于猫和小型犬心输出量的低流速的一种有前景的工具。

相似文献

2
Continuous versus bolus thermodilution cardiac output measurements--a comparative study.
Crit Care Med. 1995 May;23(5):944-9. doi: 10.1097/00003246-199505000-00025.
5
Thermodilution cardiac output--an in vitro model of low flow states.
Crit Care Med. 1986 Jan;14(1):57-9. doi: 10.1097/00003246-198601000-00013.
7
Measurement of peripheral blood flow by thermodilution techniques.
Invest Radiol. 1986 Aug;21(8):631-6. doi: 10.1097/00004424-198608000-00006.

本文引用的文献

4
Measurement of cardiac output by thermodilution in pediatric patients.
Pediatr Clin North Am. 1980 Aug;27(3):513-23. doi: 10.1016/s0031-3955(16)33889-5.
7
Measurements of variable ventricular output by thermodilution: model experiments.
J Appl Physiol. 1966 Sep;21(5):1624-32. doi: 10.1152/jappl.1966.21.5.1624.
10
Reliability of cardiac output measured by thermodilution method in small animals.
Comp Biochem Physiol. 1970 May 1;34(1):245-50. doi: 10.1016/0010-406x(70)90073-3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验