Kalyanpur A, Neklesa V P, Taylor C R, Daftary A R, Brink J A
Department of Diagnostic Radiology, Yale University School of Medicine, 2-332 SP, 333 Cedar St, New Haven, CT 06520, USA.
Radiology. 2000 Dec;217(3):772-9. doi: 10.1148/radiology.217.3.r00nv22772.
To determine acceptable levels of JPEG (Joint Photographic Experts Group) and wavelet compression for teleradiologic transmission of body computed tomographic (CT) images.
A digital test pattern (Society of Motion Picture and Television Engineers, 512 x 512 matrix) was transmitted after JPEG or wavelet compression by using point-to-point and Web-based teleradiology, respectively. Lossless, 10:1 lossy, and 20:1 lossy ratios were tested. Images were evaluated for high- and low-contrast resolution, sensitivity to small signal differences, and misregistration artifacts. Three independent observers who were blinded to the compression scheme evaluated these image quality measures in 20 clinical cases with similar levels of compression.
High-contrast resolution was not diminished with any tested level of JPEG or wavelet compression. With JPEG compression, low-contrast resolution was not lost with 10:1 lossy compression but was lost at 3% modulation with 20:1 lossy compression. With wavelet compression, there was loss of 1% modulation with 10:1 lossy compression and loss of 5% modulation with 20:1 lossy compression. Sensitivity to small signal differences (5% and 95% of the maximal signal) diminished only with 20:1 lossy wavelet compression. With 10:1 lossy compression, misregistration artifacts were mild and were equivalent with JPEG and wavelet compression. Qualitative clinical findings supported these findings.
Lossy 10:1 compression is suitable for on-call electronic transmission of body CT images as long as original images are subsequently reviewed.
确定用于身体计算机断层扫描(CT)图像远程放射学传输的JPEG(联合图像专家组)和小波压缩的可接受水平。
通过点对点和基于网络的远程放射学分别在JPEG或小波压缩后传输数字测试图案(电影与电视工程师协会,512×512矩阵)。测试了无损、10:1有损和20:1有损比率。对图像的高对比度和低对比度分辨率、对小信号差异的敏感性以及配准错误伪影进行评估。三名对压缩方案不知情的独立观察者在20例具有相似压缩水平的临床病例中评估这些图像质量指标。
在任何测试的JPEG或小波压缩水平下,高对比度分辨率均未降低。对于JPEG压缩,10:1有损压缩时低对比度分辨率未丢失,但在20:1有损压缩且调制率为3%时丢失。对于小波压缩,10:1有损压缩时调制率丢失1%,20:1有损压缩时调制率丢失5%。仅在20:1有损小波压缩时,对小信号差异(最大信号的5%和95%)的敏感性降低。在10:1有损压缩时,配准错误伪影较轻,且JPEG和小波压缩相当。定性临床结果支持这些发现。
只要随后对原始图像进行复查,10:1有损压缩适用于身体CT图像的随叫随到电子传输。