Goldstein J A, Barkin J S
University of Miami, School of Medicine/Mt. Sinai Medical Center, Division of Gastroenterology, Miami Beach, Florida 33140, USA.
Am J Gastroenterol. 2000 Dec;95(12):3423-7. doi: 10.1111/j.1572-0241.2000.03357.x.
Balloon dilation of a luminal narrowing is classically performed with a single-diameter balloon. A new multidiameter balloon that allows three progressively larger, controlled radial expansion diameters of a single-balloon catheter (controlled radial expansion, CRE) has been developed. The purpose of this study was to test the diameter consistency and dilating force at increasing pressure of the CRE balloon-dilating catheter and to compare these features with the conventional balloon-dilating catheters.
The CRE balloon catheter was compared with the Achiever, Eliminator, and Max Force using in vitro testing to determine their dilating diameters and dilating forces.
This in vitro study has shown that the new CRE balloon catheter with increasing diameter delivers consistently reproducible and progressively greater dilating force. This dilating force of the CRE balloon catheter at its third and maximum diameter inflation was similar to that of the Max Force TTS and significantly greater than that of the conventional balloon dilators, Eliminator and Achiever.
腔内狭窄的球囊扩张传统上使用单一直径的球囊进行。一种新型多直径球囊已被研发出来,它能使单球囊导管实现三个逐渐增大的、可控的径向扩张直径(可控径向扩张,CRE)。本研究的目的是测试CRE球囊扩张导管在压力增加时的直径一致性和扩张力,并将这些特性与传统球囊扩张导管进行比较。
通过体外测试将CRE球囊导管与Achiever、Eliminator和Max Force进行比较,以确定它们的扩张直径和扩张力。
这项体外研究表明,新型CRE球囊导管随着直径增加能提供一致可重复且逐渐增大的扩张力。CRE球囊导管在其第三个也是最大直径充气时的扩张力与Max Force TTS相似,且显著大于传统球囊扩张器Eliminator和Achiever。