A rotating electrode configuration is evaluated as a means to lower the detection limits of newly devised polyion-sensitive membrane electrodes (PSEs). Planar potentiometric polycation and polyanion PSEs are prepared by incorporating tridodecylmethylammonium chloride and calcium dinonylnaphthalenesulfonate, respectively, into plasticized PVC or polyurethane membranes and mounting disks of such films on an electrode body housed in a conventional rotating disk electrode apparatus. Rotation of the PSEs at 5000 rpm results in an enhancement in the detection limits toward heparin (polyanion) and protamine (polycation) of at least 1 order of magnitude (to 0.01 unit/mL for heparin; 0.02 microg/mL for protamine) over that observed when the EMF responses of the same electrodes are assessed using a stir-bar to achieve convective mass transport. A linear relationship between omega(-1/2), where omega is the rotating angular frequency, and C1/2, the polyion concentration corresponding to half the total maximum deltaEMF response toward the polyion species, is observed. It is further shown that the rotating polycation sensor can be used as an end-point detector to greatly enhance (relative to nonrotated indicator electrode) the analytical resolution and precision for measurement of low concentrations of heparin when such samples are titrated with protamine. The theoretical basis for lowering the detection limits by rotating PSEs is discussed based on the unique nonequilibrium response mechanism of such sensors.
对一种旋转电极配置进行了评估,以此作为降低新设计的聚离子敏感膜电极(PSE)检测限的一种方法。通过分别将氯化三(十二烷基)甲基铵和二壬基萘磺酸钙掺入增塑PVC或聚氨酯膜中,并将此类薄膜圆盘安装在传统旋转圆盘电极装置中的电极体上,制备了平面电位型聚阳离子和聚阴离子PSE。与使用搅拌棒实现对流质量传输来评估相同电极的电动势响应时相比,PSE以5000转/分钟的速度旋转,使得对肝素(聚阴离子)和鱼精蛋白(聚阳离子)的检测限提高了至少1个数量级(肝素达到0.01单位/毫升;鱼精蛋白达到0.02微克/毫升)。观察到ω^(-1/2)(其中ω是旋转角频率)与C^(1/2)(对应于对聚离子物种总最大Δ电动势响应一半的聚离子浓度)之间存在线性关系。进一步表明,当用鱼精蛋白滴定此类样品时,旋转聚阳离子传感器可用作终点检测器,以大大提高(相对于未旋转的指示电极)低浓度肝素测量的分析分辨率和精密度。基于此类传感器独特的非平衡响应机制,讨论了通过旋转PSE降低检测限的理论基础。