Ye L, Landen W O, Eitenmiller R R
Department of Food Science and Technology, University of Georgia, Athens 30602, USA.
J Chromatogr Sci. 2001 Jan;39(1):1-6. doi: 10.1093/chromsci/39.1.1.
A comparison of the performance of narrow-bore (2.1-mm i.d.) and standard-bore (4.6-mm i.d.) analytical silica columns having the same length is completed for the resolution of alpha-, beta-, gamma-, and delta-tocopherol. The studies are performed on high-performance liquid chromatographic equipment with minimum extracolumn contribution. Column permeabilities are 1.16 x 10(-9) and 2.48 x 10(-9) cm2 for narrow and standard bore, respectively. The narrow-bore column gives up to a 7 times increase in sensitivity compared with a standard-bore column at equivalent running times for the analytes. Approximately one-third solvent savings can be achieved with the narrow-bore column. Theoretical plates of the standard-bore column are higher than that of the narrow-bore column.