Placement of atrial pacing leads during atrial fibrillation. Feasibility and subsequent lead performance.

作者信息

Marshall H J, Griffith M J, Gammage M D

机构信息

University Department of Cardiovascular Medicine, Queen Elizabeth Hospital, Birmingham, U.K.

出版信息

Europace. 1999 Apr;1(2):131-4. doi: 10.1053/eupc.1998.0030.

Abstract

AIMS

To assess the feasibility of placing permanent atrial pacing leads during atrial fibrillation (AF) and whether such leads function satisfactorily.

METHODS AND RESULTS

Prospective study of 17 consecutive patients in whom permanent atrial leads were positioned during an episode of paroxysmal AF. Fluoroscopic position ('figure of 8' or side-to-side movement and anterior position in RAO projection), lead impedance (> 300 but < 1000 ohms) and intracardiac electrogram (average peak to peak amplitude > 1 mV) were used to define an acceptable lead position. At 8 weeks post implant we measured: pulse duration pacing threshold at 5 V; lead impedance at 5 V and 0.5 ms; intracardiac electrogram (EGM) signal amplitude. At the end of the study we reviewed patients to establish whether AF had become permanent. In all patients, follow-up demonstrated satisfactory lead function. All leads had impedances between 300 and 1000 ohms. Pacing thresholds were all < 0.1 ms at 5 V. Mean atrial EGM amplitude seen in sinus rhythm was 3.3 mV (range 1.2-8.4); in patients where all follow-up was in AF in was 2.1 mV (range 1.5-2.5). Nine patients (53%) developed permanent AF.

CONCLUSION

Placing atrial leads during AF is feasible using the technique described. However, some patients progress to chronic AF, eliminating the benefits of atrial pacing.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验