Rowlands J A, Taylor K W
Med Phys. 1984 Sep-Oct;11(5):597-601. doi: 10.1118/1.595556.
For monoenergetic x rays, the detective quantum efficiency (DQE) of an x-ray image intensifier (XRII) obtained by measuring the ratio of signal to root-mean-square (rms) noise is compared with the DQE obtained by the scintillation spectrum method. To obtain agreement between the two methods, the rms noise must be measured at a very low frequency because of the temporal response RXRII (f) of the XRII. This takes a long time and the measurement of the DQE from the scintillation spectrum is much quicker. Measurement of RXRII (f) shows that it is dependent on x-ray flux rate. RXRII (f) measured at the same flux rate as DQE permits correction of DQE measurements to zero frequency. In order to estimate DQE for polyenergetic beams, we need to know as a function of energy not only the DQE but the amount of energy deposited per absorbed x ray. The latter is also obtained from the rms measurement.
对于单能X射线,通过测量信号与均方根(rms)噪声之比获得的X射线图像增强器(XRII)的探测量子效率(DQE),与通过闪烁光谱法获得的DQE进行了比较。为了使两种方法达成一致,由于XRII的时间响应RXRII(f),必须在非常低的频率下测量rms噪声。这需要很长时间,而从闪烁光谱测量DQE则要快得多。对RXRII(f)的测量表明,它取决于X射线通量率。在与DQE相同的通量率下测量RXRII(f),可以将DQE测量值校正到零频率。为了估计多能束的DQE,我们不仅需要知道作为能量函数的DQE,还需要知道每次吸收的X射线沉积的能量。后者也从rms测量中获得。