Suppr超能文献

在线Sensicath血气监测系统的临床评估

Clinical evaluation of the on-line Sensicath blood gas monitoring system.

作者信息

Myklejord D J, Pritzker M R, Nicoloff D M, Emery A M, Emery R W

机构信息

Minneapolis Heart Institute, Minneapolis, MN 55407, USA.

出版信息

Heart Surg Forum. 1998;1(1):60-4.

Abstract

BACKGROUND

Blood gas analysis is an integral part of the management of open heart surgery and post-surgical intensive care patients. This study was conducted to compare the results of on-line blood gas analysis using the new Sensicath (SC) optical sensor technology against standard blood gas assay.

METHODS

Comparative blood gas analysis was performed on 45 postoperative cardiovascular surgical patients and 5 patients during cardiopulmonary bypass. Simultaneous samples were obtained. One sample was sent to the central hospital laboratory for measurement using a Radiometer ABL 500 blood gas analyzer and the second sample was drawn past the optical sensor. Comparisons between the two methods for pH, pO2 and pCO2 were analyzed using linear regression and the method of Bland and Altman for the intensive care unit samples (N = 451) while the T test was used to compare data points obtained during cardiopulmonary bypass (N = 57). Bias, accuracy and precision were also calculated.

RESULTS

The regression lines (r2) for pH, pO2, and pCO2 and were 0.69, 0.94, and 0.68 with slopes of 1.000, 0.0957, and 1.052 respectively for postoperative surgical patients. During cardiopulmonary bypass, the T test did not show any significant difference between the Sensicath and the reference values for the arterial and venous pH, pO2 and pCO2.

CONCLUSIONS

In summary, this study found the Sensicath to be an accurate, useful tool which permits immediate blood gas analysis without blood exposure and constitutes an advance in the field of intensive care monitoring.

摘要

背景

血气分析是心脏直视手术及术后重症监护患者管理的一个重要组成部分。本研究旨在比较使用新型 Sensicath(SC)光学传感器技术进行在线血气分析与标准血气检测的结果。

方法

对 45 例心血管外科术后患者及 5 例体外循环患者进行了对比血气分析。同时采集样本。一份样本送至中心医院实验室,使用 Radiometer ABL 500 血气分析仪进行检测,另一份样本通过光学传感器抽取。对于重症监护病房的样本(N = 451),使用线性回归以及 Bland 和 Altman 方法分析两种方法在 pH、pO2 和 pCO2 方面的差异,而对于体外循环期间获得的数据点(N = 57),使用 T 检验进行比较。还计算了偏差、准确性和精密度。

结果

术后患者的 pH、pO2 和 pCO2 的回归线(r2)分别为 0.69、0.94 和 0.68,斜率分别为 1.000、0.0957 和 1.052。在体外循环期间,T 检验未显示 Sensicath 与动脉和静脉 pH、pO2 和 pCO2 的参考值之间存在任何显著差异。

结论

总之,本研究发现 Sensicath 是一种准确、有用的工具,可在不接触血液的情况下即时进行血气分析,是重症监护监测领域的一项进步。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验