Intracellular calcium handling plays an important role in cardiac electrophysiology. Using two fluorescent indicators, we developed an optical mapping system that is capable of measuring calcium transients and action potentials at 256 recording sites simultaneously from the intact guinea pig heart. On the basis of in vitro measurements of dye excitation and emission spectra, excitation and emission filters at 515 +/- 5 and >695 nm, respectively, were used to measure action potentials with di-4-ANEPPS, and excitation and emission filters at 365 +/- 25 and 485 +/- 5 nm, respectively, were used to measure calcium transients with indo 1. The percent error due to spectral overlap was small when action potentials were measured (1.7 +/- 1.0%, n = 3) and negligible when calcium transients were measured (0%, n = 3). Recordings of calcium transients, action potentials, and isochrone maps of depolarization time and the time of calcium transient onset indicated negligible error due to fluorescence emission overlap. These data demonstrate that the error due to spectral overlap of indo 1 and di-4-ANEPPS is sufficiently small, such that optical mapping techniques can be used to measure calcium transients and action potentials simultaneously in the intact heart.
细胞内钙处理在心脏电生理学中起着重要作用。我们使用两种荧光指示剂开发了一种光学映射系统,该系统能够同时在256个记录位点测量来自完整豚鼠心脏的钙瞬变和动作电位。基于染料激发和发射光谱的体外测量,分别使用515±5和>695nm的激发和发射滤光片,用di-4-ANEPPS测量动作电位,分别使用365±25和485±5nm的激发和发射滤光片,用indo 1测量钙瞬变。测量动作电位时,由于光谱重叠引起的百分比误差较小(1.7±1.0%,n = 3),测量钙瞬变时可忽略不计(0%,n = 3)。钙瞬变、动作电位以及去极化时间和钙瞬变起始时间的等时线图的记录表明,由于荧光发射重叠引起的误差可忽略不计。这些数据表明,indo 1和di-4-ANEPPS光谱重叠引起的误差足够小,因此光学映射技术可用于在完整心脏中同时测量钙瞬变和动作电位。