Suppr超能文献

Determination of moxifloxacin in growth media by high-performance liquid chromatography.

作者信息

Ba B B, Etienne R, Ducint D, Quentin C, Saux M C

机构信息

Laboratoire de Pharmacocinétique et de Pharmacie Clinique, Faculté de Pharmacie, Université Victor Segalen Bordeaux 2, France.

出版信息

J Chromatogr B Biomed Sci Appl. 2001 Apr 15;754(1):107-12. doi: 10.1016/s0378-4347(00)00591-0.

Abstract

A direct injection high-performance liquid chromatographic method with column switching has been developed to determine moxifloxacin in Mueller-Hinton broth. A LiChrocart 4-4 pre-column filled with a LiChrospher 100 RP 18, 5 microm and a 150 x 4.6 mm I.D. column packed with a Supelcozil ABZ+ Plus were used and led to a retention time of 5.70 min. Fluorescence detection allowed one to reach a quantification limit of 0.05 microg/ml with a 100-microl sample size. The standard curves were linear from 0.05 to 3.2 microg/ml. Intra- and inter-day imprecisions within the linearity range were < or =4.76 and < or =5.75%, respectively. The mean relative errors for the same day and the day-to-day inaccuracies ranged from -2.93 to +4.50% and from -1.10 to +6.00%, respectively. The method was demonstrated to be useful for pharmacokinetic-pharmacodynamic studies of moxifloxacin in an in vitro model.