Suppr超能文献

正压通气机中的漏气补偿:一项肺模型研究。

Leak compensation in positive pressure ventilators: a lung model study.

作者信息

Mehta S, McCool F D, Hill N S

机构信息

Mt Sinai Hospital, University of Toronto, Canada.

出版信息

Eur Respir J. 2001 Feb;17(2):259-67. doi: 10.1183/09031936.01.17202590.

Abstract

Leak compensating abilities of six different positive pressure ventilators commonly used to deliver noninvasive positive pressure ventilation, including the bilevel positive airway pressure (BiPAP) S/T-D and Quantum (Respironics Inc, Murrysville, PA, USA), 335 and O'NYX (Mallinckrodt Inc, St Louis, MO, USA), PLV 102 (Respironics), and Siemens Servo 900C (Siemens Inc, Danvers, MA, USA). Using a test lung model, compensatory capabilities of the ventilators were tested for smaller and larger leaks using the assist/control or timed modes. Back-up rate was 20 min(-1), inspiratory pressure was 18 cmH2O, and expiratory pressure was 5 cmH2O. It was found that even in the absence of air leaking, delivered tidal volume differed substantially between the ventilators during use of pressure-targeted modes, depending on inspiratory flows, inaccuracies in set versus delivered pressures, and inspiratory duration. Also during pressure-targeted ventilation, increasing the tI/ttot up to, but not beyond, 0.5 improved compensation by lengthening inspiratory duration, whereas use of a sensitive flow trigger setting tended to cause autocycling during leaking, interfering with compensation. Leaking interfered with cycling of the BiPAP S/T, inverting the I:E ratio, shortening expiratory time, and reducing delivered tidal volume. Volume-targeted modes achieved limited compensation for small air leaks, but compensated poorly for large leaks. To conclude, leak-compensating capabilities differ markedly between ventilators but pressure-targeted ventilators are preferred for noninvasive positive pressure ventilation in patients with substantial air leaking. Adequate inspiratory flows and durations should be used, triggering sensitivity should be adjusted to prevent autocycling, and a mechanism should be available to limit inspiratory time and avoid I:E ratio inversion.

摘要

六种常用于提供无创正压通气的不同正压通气机的漏气补偿能力,包括双水平气道正压(BiPAP)S/T-D和Quantum(美国宾夕法尼亚州默里斯维尔的瑞思迈公司)、335和O'NYX(美国密苏里州圣路易斯的马利克若德公司)、PLV 102(瑞思迈公司)以及西门子Servo 900C(美国马萨诸塞州丹弗斯的西门子公司)。使用测试肺模型,在辅助/控制或定时模式下,针对较小和较大漏气情况测试了通气机的补偿能力。后备频率为20次/分钟,吸气压力为18 cmH₂O,呼气压力为5 cmH₂O。研究发现,即使在没有空气泄漏的情况下,在使用压力目标模式时,不同通气机输送的潮气量也存在显著差异,这取决于吸气流量、设定压力与输送压力的误差以及吸气持续时间。同样在压力目标通气期间,将吸气时间与总周期时间(tI/ttot)增加至但不超过0.5可通过延长吸气持续时间改善补偿,而使用敏感的流量触发设置在漏气期间往往会导致自动切换,干扰补偿。漏气干扰了BiPAP S/T的切换,使吸气与呼气时间比(I:E)倒置,缩短呼气时间,并减少输送的潮气量。容量目标模式对小量空气泄漏的补偿有限,但对大量泄漏的补偿较差。总之,不同通气机的漏气补偿能力存在显著差异,但对于有大量空气泄漏的患者,无创正压通气首选压力目标通气机。应使用足够的吸气流量和持续时间,调整触发敏感度以防止自动切换,并且应有一种机制来限制吸气时间并避免I:E比倒置。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验