Pressure control ventilation: three anesthesia ventilators compared using an infant lung model.

作者信息

Stayer S A, Bent S T, Skjonsby B S, Frolov A, Andropoulos D B

机构信息

Department of Anesthesiology, Baylor College of Medicine, Texas Children's Hospital, Houston, Texas 77030, USA.

出版信息

Anesth Analg. 2000 Nov;91(5):1145-50. doi: 10.1097/00000539-200011000-00020.

Abstract

UNLABELLED

We compared three ventilators-Servo 900C (Siemens Medical Systems, Danvers, MA), Aestiva 3000 (Datex-Ohmeda, Madison, WI), and NAD 6000 (North American Dräger, Telford, PA)-set to deliver pressure control ventilation using an infant test lung model. Ventilator settings were selected to test "near-maximum" settings that would be used for a neonatal patient (peak inspiratory pressure [PIP] 30 cm H(2)O) or older child (PIP 60 cm H(2)O). When adjusted for set inspiratory pressure and compliance, the average tidal volume (V(t)) produced by the NAD 6000 was 5.8 mL less than the Servo 900C (P: = 0. 103), and the average V(t) produced by the Aestiva 3000 was 18.9 mL less than the Servo 900C (P: < 0.001). The Servo 900C generated increased peak pressures, tending to overshoot the set maximum inflating pressures, especially during rapid respiratory rates with decreased inspiratory times. The Aestiva 3000 did not achieve the set PIP during testing conditions of decreased inspiratory times, and the NAD 6000 was not greatly affected by changes in inspiratory time. All three ventilators measured expiratory V(t) to be larger than the actual V(t) delivered to the lung; however, the NAD 6000 was more accurate.

IMPLICATIONS

There are differences in performance of ventilators when set to deliver pressure control ventilation to an infant test lung model.

摘要

未标注

我们使用婴儿测试肺模型,比较了三种设置为进行压力控制通气的呼吸机——Servo 900C(西门子医疗系统公司,马萨诸塞州丹弗斯)、Aestiva 3000(Datex - Ohmeda公司,威斯康星州麦迪逊)和NAD 6000(北美德尔格公司,宾夕法尼亚州特尔福德)。选择呼吸机设置以测试用于新生儿患者(吸气峰压[PIP] 30 cm H₂O)或大龄儿童(PIP 60 cm H₂O)的“接近最大”设置。当根据设定的吸气压力和顺应性进行调整时,NAD 6000产生的平均潮气量(Vₜ)比Servo 900C少5.8 mL(P = 0.103),Aestiva 3000产生的平均Vₜ比Servo 900C少18.9 mL(P < 0.001)。Servo 900C产生的峰压增加,倾向于超过设定的最大充气压力,尤其是在呼吸频率快且吸气时间缩短时。在吸气时间缩短的测试条件下,Aestiva 3000未达到设定的PIP,而NAD 6000受吸气时间变化的影响不大。所有三种呼吸机测得的呼气Vₜ均大于输送到肺的实际Vₜ;然而,NAD 6000更准确。

启示

当设置为向婴儿测试肺模型进行压力控制通气时,呼吸机的性能存在差异。