评估不同记录参数以建立啮齿动物闪光视网膜电图标准。

Evaluation of different recording parameters to establish a standard for flash electroretinography in rodents.

作者信息

Bayer A U, Cook P, Brodie S E, Maag K P, Mittag T

机构信息

Department of Ophthalmology, Box 1183, The Mount Sinai Medical Center, Mount Sinai School of Medicine of New York University, 1 Gustave L. Levy Place, New York, NY 10029, USA.

出版信息

Vision Res. 2001 Aug;41(17):2173-85. doi: 10.1016/s0042-6989(01)00103-1.

Abstract

Different electrodes and stimulus protocols commonly used for electroretinography in rodent eyes were compared for convenience of use, degree of damage to corneal epithelium, and for magnitude of amplitude, reproducibility, left versus right eye accuracy, and reliability of recorded parameters of the flash electroretinogram (ERG). Adult C57BL/6 pigmented mice and albino Wistar rats were used to determine scotopic ERGs in response to Ganzfeld or strobe-light stimulation and light-adapted (photopic) ERGs recorded from both eyes at the same time. Test-retest data were used for statistical analyses to compare a monopolar gold-wire contact lens electrode (CLE), a cotton-wick silver-silver chloride electrode (CSCE), a DTL fiber electrode (DTLE), and a circular stainless steel wire electrode (SSE). Corneas were evaluated for abrasion after ERG recordings using fluorescein staining and also for the time taken, ease of insertion, and re-insertions required for the different electrodes. Compared to CSCE, DTLE, and SSE, the ERG potentials recorded by CLE had significantly larger scotopic amplitudes and oscillatory potentials under strobe or Ganzfeld stimulation and for light-adapted ERG b-wave amplitudes in both mice and rats. In analyzing test-retest data of scotopic ERG a-wave and b-wave amplitudes, the intraclass correlation coefficient showed the best agreement for the CLE (range 0.61-0.94) compared to the SSE (0.13-0.77), DTLE (0.02-0.69), and CSCE (0.12-0.51). In mice and rats, logistic regression analyses revealed significant correlations for amplitudes of most scotopic ERG parameters between contralateral eyes obtained with CLE and for some ERG components recorded by SSE. When comparing ERG amplitudes for stimulation by strobe or Ganzfeld, the difference was least with the CLE compared to DTLE, CSCE, or SSE. The time taken to insert the four different electrodes was greatest for the CLE in both mice and rats. The extent of corneal abrasion resulting from electrode use in mice was largest for the SSE followed by the CLE. However, in rats there was almost no corneal damage after ERG recordings with the CLE. Because of the stability of eye contact, the CLE allows ERGs to be determined over a longer recording session. Recording of scotopic and photopic (light-adapted) ERGs in rodents with monopolar gold-wire contact lens electrodes provides greater amplitudes and higher reproducibility when compared to other commonly used corneal electrodes. These electrodes are significantly better overall than others that were evaluated and should be considered for a standard protocol to monitor retinal function in rodent eyes.

摘要

比较了啮齿动物眼睛视网膜电图常用的不同电极和刺激方案,评估其使用便利性、对角膜上皮的损伤程度以及闪光视网膜电图(ERG)的振幅大小、可重复性、左右眼准确性和记录参数的可靠性。使用成年C57BL/6有色小鼠和白化Wistar大鼠来测定对全视野或频闪刺激的暗视ERG,并同时记录双眼的明视(光适应)ERG。采用重测数据进行统计分析,比较单极金线隐形眼镜电极(CLE)、棉芯银 - 氯化银电极(CSCE)、DTL纤维电极(DTLE)和圆形不锈钢丝电极(SSE)。ERG记录后,使用荧光素染色评估角膜磨损情况,并记录不同电极插入所需时间、插入难易程度和重新插入次数。与CSCE、DTLE和SSE相比,CLE记录的ERG电位在频闪或全视野刺激下以及小鼠和大鼠的明视ERG b波振幅方面,暗视振幅和振荡电位明显更大。在分析暗视ERG a波和b波振幅的重测数据时,组内相关系数显示CLE(范围0.61 - 0.94)与SSE(0.13 - 0.77)、DTLE(0.02 - 0.69)和CSCE(0.12 - 0.51)相比一致性最佳。在小鼠和大鼠中,逻辑回归分析显示,使用CLE获得的对侧眼之间大多数暗视ERG参数的振幅以及SSE记录的一些ERG成分之间存在显著相关性。当比较频闪或全视野刺激的ERG振幅时,与DTLE、CSCE或SSE相比,CLE的差异最小。在小鼠和大鼠中,插入四种不同电极所需时间CLE最长。小鼠中因电极使用导致的角膜磨损程度SSE最大,其次是CLE。然而,在大鼠中,使用CLE进行ERG记录后几乎没有角膜损伤。由于眼睛接触的稳定性,CLE允许在更长的记录时间内测定ERG。与其他常用的角膜电极相比,使用单极金线隐形眼镜电极记录啮齿动物的暗视和明视(光适应)ERG可提供更大的振幅和更高的可重复性。这些电极总体上明显优于其他评估过的电极,应考虑将其作为监测啮齿动物眼睛视网膜功能的标准方案。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验