Suppr超能文献

接触镜和粘性皮肤电极的 ERG 的定性和定量比较。

Qualitative and quantitative comparison of ERGs with contact lens and adhesive skin electrodes.

机构信息

Department of Ophthalmology, NYU Langone Health, New York, NY, USA.

出版信息

Doc Ophthalmol. 2022 Jun;144(3):203-215. doi: 10.1007/s10633-022-09868-w. Epub 2022 Mar 18.

Abstract

PURPOSE

Traditional ERGs recorded using corneal electrodes can be difficult for some patients to tolerate. In the last several years, adhesive skin electrodes have gained in acceptance. In this report we present a qualitative comparison of waveforms as well as a quantitative analysis of correlation of amplitudes and implicit times of simultaneous ERG recordings using contact lens and skin electrodes.

METHODS

89 subjects were included; all were referred for full-field ERG testing for multiple indications. ERGs (obtained according to ISCEV standards) were recorded simultaneously from both eyes with ERG-jet corneal contact lens electrodes and LKC Technologies Sensor Strip adhesive skin electrodes using multi-channel instrumentation (Diagnosys LLC, Espion3). Waveforms, a-wave and b-wave amplitudes and implicit times were compared.

RESULTS

Waveform morphologies were similar between electrode types. Regression coefficients (conversion factors) for a-wave and b-wave amplitudes under both photopic and scotopic conditions were tightly clustered. Regression coefficients for implicit times were nearly equal to 1.0. The regression coefficient for the entire amplitude dataset was 0.349, with an overall correlation of 0. 869 between amplitude recorded with skin and contact lens electrodes. The regression coefficient for the entire implicit time dataset was 0.967, with an overall correlation of 0.964 between skin and contact lens electrodes.

CONCLUSIONS

Our best estimate for the conversion factor between ERG amplitudes recorded with adhesive skin electrodes and contact lens electrodes is 0.349-amplitudes with skin electrodes are about 1/3 the amplitudes recorded simultaneously from the same eyes with contact lens electrodes, with a high correlation. Implicit times are nearly identical for the two electrode types.

摘要

目的

使用角膜电极记录传统的 ERG 可能会让一些患者难以忍受。在过去的几年中,粘性皮肤电极的应用越来越受到认可。在本报告中,我们对使用接触镜和皮肤电极同时记录的波形进行了定性比较,并对振幅和潜伏期的相关性进行了定量分析。

方法

纳入 89 例受试者;所有受试者均因多种适应症接受全视野 ERG 检查。使用多通道仪器(Diagnosys LLC,Espion3),根据 ISCEV 标准,同时从双眼记录 ERG-jet 角膜接触镜电极和 LKC Technologies Sensor Strip 粘性皮肤电极的 ERG(获得)。比较了波形、a 波和 b 波振幅和潜伏期。

结果

两种电极类型的波形形态相似。明视和暗视条件下 a 波和 b 波振幅的回归系数(转换因子)紧密聚集。潜伏期的回归系数接近 1.0。整个振幅数据集的回归系数为 0.349,皮肤和接触镜电极记录的振幅之间的总相关系数为 0.869。整个潜伏期数据集的回归系数为 0.967,皮肤和接触镜电极之间的总相关系数为 0.964。

结论

我们对使用粘性皮肤电极和接触镜电极记录的 ERG 振幅之间的转换因子的最佳估计值为 0.349-皮肤电极记录的振幅约为同时使用接触镜电极从同一眼睛记录的振幅的 1/3,具有较高的相关性。两种电极类型的潜伏期几乎相同。