The reliability of pacemaker electrode-leads.

作者信息

Green G D

出版信息

J Electrocardiol. 1975;8(2):195-200. doi: 10.1016/s0022-0736(75)80030-6.

Abstract

Extensive experience has been obtained with Medtronic pacemakers, in Glasgow where over 400 patients have received their primary implant. In this review, which covers a period of seven and a half years ending December 31, 1973, the technical reliability of electrode-leads is considered in terms of failed implant lifetimes and incomplete implant lifetimes. Too few myocardial electrode-leads have been used to reach firm conclusions about its reliability, but the incidence of failures with the endocardial bipolar electrode-lead (Type 5818) has so far been small and it is evidently more reliable than its predecessor (Type 5816). More time must elapse, however, before it can be claimed with confidence that the present design of bipolar catheter will match the theoretical implant lifetimes of generators with the long-life power sources. Insulation failures are usually of no immediate clinical significance whereas any broken conductor is hazardous. The use of multistranded conductors is therefore advocated. All pacemakers should be retrieved after the deaths of patients.

摘要

在格拉斯哥,已有400多名患者首次植入美敦力起搏器,积累了丰富的经验。本综述涵盖截至1973年12月31日的七年半时间,从植入失败的寿命和未完成植入的寿命方面考虑电极导线的技术可靠性。用于心肌的电极导线数量过少,无法就其可靠性得出确凿结论,但心内膜双极电极导线(5818型)的故障发生率到目前为止一直很低,显然比其前身(5816型)更可靠。然而,在能够自信地宣称目前双极导管的设计将与带有长寿命电源的发生器的理论植入寿命相匹配之前,还需要更长的时间。绝缘故障通常没有直接的临床意义,而任何导线断裂都是危险的。因此提倡使用多股导线。所有起搏器都应在患者死亡后取出。