Wirtz S, Schulte H D, Winter J, Godehardt E, Kunert J
Department of Thoracic and Cardiovascular Surgery, Heinrich Heine University of Düsseldorf, FRG.
Thorac Cardiovasc Surg. 1989 Jun;37(3):163-8. doi: 10.1055/s-2007-1020310.
Temporary myocardial pacing leads are routinely used for diagnosis and treatment of postoperative arrhythmias following open-heart surgery. The intention of this study is to compare five different electrodes for reliability during the postoperative period. A standardised technique of implantation was used to place 147 ventricular and 81 atrial wire electrodes in 149 patients. During operation and then daily over the next ten days, the stimulation threshold, P- and R-wave sensing, and impedance were measured with a pacing system analyser to evaluate the reliability of the pacing and sensing function. Five measurements were taken at each time; the mean values were statistically analysed. The number of failures in pacing and sensing and the course of the tested parameters during the postoperative period were compared. As a result, bipolar pacing leads (Osypka TME 64a/66a) proved to be more reliable than unipolar wires (Ethicon HD 12S, Osypka TME 60/60a), particularly in atrial sensing. Also the Medtronic 6400/6500 showed good reliability for the ventricle, so that they can be recommended for VVI-mode pacing. In conclusion, for reliable function during the postoperative course up to 10 days a specially developed bipolar temporary pacemaker electrode can be recommended for both atrial and ventricular or sequential pacing.
临时心肌起搏导线常用于心脏直视手术后心律失常的诊断和治疗。本研究旨在比较五种不同电极在术后期间的可靠性。采用标准化植入技术,在149例患者中放置了147根心室和81根心房导线电极。在手术期间以及术后接下来的十天每天使用起搏系统分析仪测量刺激阈值、P波和R波感知以及阻抗,以评估起搏和感知功能的可靠性。每次进行五次测量;对平均值进行统计分析。比较了术后期间起搏和感知失败的次数以及测试参数的变化过程。结果表明,双极起搏导线(Osypka TME 64a/66a)比单极导线(Ethicon HD 12S,Osypka TME 60/60a)更可靠,尤其是在心房感知方面。美敦力6400/6500对心室也显示出良好的可靠性,因此可推荐用于VVI模式起搏。总之,为了在术后长达10天的过程中实现可靠功能,对于心房、心室或顺序起搏,可推荐一种专门开发的双极临时起搏器电极。