Strohm O, Kivelitz D, Gross W, Schulz-Menger J, Liu X, Hamm B, Dietz R, Friedrich M G
Universitätsklinikum, Medizinische Fakultät der Humboldt-Universität zu Berlin, Medizinische Klinik mit Schwerpunkt Molekulare und Klinische Kardiologie (Buch), Berlin, Germany.
J Cardiovasc Magn Reson. 1999;1(3):239-45. doi: 10.3109/10976649909088336.
The safety of most available implantable intracoronary stents during magnetic resonance imaging (MRI) has not been sufficiently tested. Minor, but possibly clinically significant, increases in temperature have not been excluded. We measured temperature changes of 14 different stents clinically in use or currently tested for release on the world market. Stents were examined in 1.0- and 1.5-T MR scanners with multiple sequences used in routine cardiac and thoracic MRI examinations ("clinical worst case") and after implantation of the stents into the coronary arteries of excised pig hearts (1.5-T scanner only). We used a highly sensitive infrared camera with a thermal resolution of 10 mK and did not see significant heating of any stent during the examinations. We conclude that MRI is safe in patients with the currently available intracoronary stents.