Ma Ying, Huang Bill X, Channing Michael A, Eckelman William C
PET Department, Warren G. Magnuson Clinical Center, National Institutes of Health, Building 10, Room 1C 401, Bethesda, MD 20892, USA.
Nucl Med Biol. 2002 Jan;29(1):125-9. doi: 10.1016/s0969-8051(01)00269-4.
A high-performance liquid chromatography-tandem mass spectrometric method (LC/MS/MS) was developed and validated for the quantitative analysis of Kryptofix (K-222) in the radiopharmaceuticals of 2-deoxy-2-[(18)F] fluoro-D-glucose (2-[(18)F]FDG) and 3-(3-((3-fluoropropyl)thio)-1,2,5-thiadiazol-4-yl)-1,2,5,6-tetrahydro-1-methylpyridine (FPTZTP). With an internal standard, the limit of quantitation for K-222 was 1.0 ng/ml. This is so far the most sensitive method for the quantification of K-222. Excellent linearity (RSQ = 0.9997) was obtained over the range of 1.0-100 ng/ml. Good precision and accuracy were also observed. The method is amenable to the validation of radiosynthetic methods.