Suppr超能文献

Homogenous enzyme immunoassay for cyclosporine in whole blood using the EMIT 2000 cyclosporine specific assay with the COBAS MIRA-plus analyzer.

作者信息

Kimura S, Iyama S, Yamaguchi Y, Kanakura Y

机构信息

Laboratory for Clinical Investigation, Osaka University Hospital, Suita, Osaka, Japan.

出版信息

J Clin Lab Anal. 2001;15(6):319-23. doi: 10.1002/jcla.1044.

Abstract

We describe the evaluation of the EMIT 2000 cyclosporine specific assay kit, an enzyme-multiplied immunoassay for cyclosporine in whole blood, with a COBAS MIRA-plus analyzer. The enzyme used for the assay was glucose-6-phosphate dehydrogenase (EC 1.1.1.49 G6PDH) from Leuconostoc mesenteroides; the monoclonal antibody is fairly specific for cyclosporine, and is not reactive with most metabolites. The assay principle is based on competitive immunoassay with G6PDH-labeled cyclosporine and cyclosporine in sample to the anticyclosporine mouse monoclonal antibody binding site. The within-assay coefficient of variation (CV) of this method was 2.7-4.2% (n = 10) at the levels of 56.2-339.7 microg/L. Day-to-day CVs ranged from 4.2-8.1% at the levels of 47.2-350.2 microg/L. The within-day CVs ranged from 2.0-6.4% at the levels of 43.3-330.5 microg/L. The functional detection limit was 24.9 microg/L. Samples treated with pretreatment reagent were stable at least 5 hr. Calibration was stable at least 10 days. The analytical recovery was 81-109%. The correlation between values obtained with the EMIT 2000 cyclosporine specific assay kit (y) and fluorescence polarization immunoassay (FPIA) (TDxFLx) (x) was: y = 0.880x - 13.053 microg/L (r = 0.984, Sy/x = 15.968, n = 71) with a mean difference of 31.42 +/- 19.89 microg/L ((TDxFLx - EMIT 2000) +/- SD); for the FPIA (AxSYM) (x): y = 0.989 - 4.144 microg/L (r = 0.981, Sy/x = 17.478, n = 71) with a mean difference of 5.56 +/- 17.38 microg/L ((AxSYM - EMIT 2000) +/- SD); and for the radioimmunoassay (RIA, CYCLO-Trac SP) (x): y = 0.893 - 6.764 microg/L (r = 0.993, Sy/x = 10.582, n = 71) with a mean difference of 22.18 +/- 14.98 microg/L ((RIA - EMIT 2000) +/- SD) using the Bland-Altman technique.

摘要

相似文献

3
Progress report of an external quality assessment scheme for cyclosporine assay.
Ther Drug Monit. 1996 Jun;18(3):273-9. doi: 10.1097/00007691-199606000-00009.
9
EMIT cyclosporine assay: development of an application protocol for Technicon AXON System.
Clin Biochem. 1994 Oct;27(5):387-94. doi: 10.1016/0009-9120(94)90043-4.
10
Evaluation of the AxSYM monoclonal cyclosporin assay and comparison with radioimmunoassay.
Ther Drug Monit. 2000 Aug;22(4):474-80. doi: 10.1097/00007691-200008000-00018.

本文引用的文献

6
Cyclosporine disposition and metabolite profiles in renal transplant patients receiving a microemulsion formulation.
Ther Drug Monit. 1994 Oct;16(5):519-25. doi: 10.1097/00007691-199410000-00014.
7
Cyclosporin A. Clinical pharmacology and therapeutic potential.
Drugs. 1982 Oct;24(4):322-34. doi: 10.2165/00003495-198224040-00004.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验