Modeling bipolar phase-shifted multielectrode catheter ablation.

作者信息

Tungjitkusolmun Supan, Haemmerich Dieter, Cao Hong, Tsai Jang-zern, Choy Young Bin, Vorperian Vicken R, Webster John G

机构信息

Department of Electronics Engineering, King Mongkut's Institute of Technology Ladkrabang, Bangkok, Thailand.

出版信息

IEEE Trans Biomed Eng. 2002 Jan;49(1):10-7. doi: 10.1109/10.972835.

Abstract

Atrial fibrillation (AFIB) is a common clinical problem affecting approximately 0.5-1% of the United States population. Radio-frequency (RF) multielectrode catheter (MEC) ablation has successes in curing AFIB. We utilized finite-element method analysis to determine the myocardial temperature distribution after 30 s, 80 degrees C temperature-controlled unipolar ablation using three 7F 12.5-mm electrodes with 2-mm interelectrode spacing MEC. Numerical results demonstrated that cold spots occurred at the edges of the middle electrode and hot spots at the side electrodes. We introduced the bipolar phase-shifted technique for RF energy delivery of MEC ablation. We determined the optimal phase-shift (phi) between the two sinusoidal voltage sources of a simplified two-dimensional finite-element model. At the optimal phi, we can achieve a temperature distribution that minimizes the difference between temperatures at electrode edges. We also studied the effects of myocardial electric conductivity (sigma), thermal conductivity (k), and the electrode spacing on the optimal phi. When we varied sigma and kappa from 50% to 150%, optimal phi ranged from 29.5 degrees to 23.5 degrees, and in the vicinity of 26.5 degrees, respectively. The optimal phi for 3-mm spacing MEC was 30.5 degrees. We show the design of a simplified bipolar phase-shifted MEC ablation system.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验