Den Boer Jacques A, Bourland Joe D, Nyenhuis John A, Ham Cornelis L G, Engels Johan M L, Hebrank Franz X, Frese Georg, Schaefer Daniel J
J Magn Reson Imaging. 2002 May;15(5):520-5. doi: 10.1002/jmri.10110.
To compare thresholds for peripheral nerve stimulation from gradient switching in whole body magnetic resonance (MR) equipment of different design.
Threshold data obtained in three experiments were reformatted into a single joint data set describing thresholds for anterio-posterior (AP) gradient orientation and Echo Planar Imaging (EPI) waveforms with bipolar ramp times between 0.07 and 1.2 ms. Reformatting included the use of: a) the rate of change of the maximum field in the patient space as a measure of gradient output, b) lowest observable thresholds, c) lognormal distribution of thresholds, and d) equal standard deviation (SD) of all samples.
The joint data fit a hyperbolic threshold function. The residues were not significantly different between experiments.
Then expressed in appropriate format, the thresholds for peripheral nerve stimulation in volunteers for whole body MR equipment can be described with a hyperbolic threshold curve with rheobase 18.8 +/- 0.6 Tesla/second and chronaxie 0.36 +/- 0.02 milliseconds.
比较不同设计的全身磁共振(MR)设备中梯度切换引起的外周神经刺激阈值。
将在三个实验中获得的阈值数据重新格式化为一个联合数据集,该数据集描述了前后(AP)梯度方向和回波平面成像(EPI)波形的阈值,双极斜坡时间在0.07至1.2毫秒之间。重新格式化包括使用:a)患者空间中最大场的变化率作为梯度输出的度量,b)最低可观察阈值,c)阈值的对数正态分布,以及d)所有样本的相等标准差(SD)。
联合数据符合双曲线阈值函数。实验之间的残差无显著差异。
以适当格式表示时,全身MR设备志愿者外周神经刺激的阈值可用具有基强度18.8±0.6特斯拉/秒和时值0.36±0.02毫秒的双曲线阈值曲线来描述。