不同骨质量下牙种植体稳定性的共振频率评估:一种数值方法。

Resonance frequency assessment of dental implant stability with various bone qualities: a numerical approach.

作者信息

Huang Haw-Ming, Lee Sheng-Yang, Yeh Ching-Ying, Lin Che-Tong

机构信息

School of Medical Technology, Taipei Medical University, Taiwan.

出版信息

Clin Oral Implants Res. 2002 Feb;13(1):65-74. doi: 10.1034/j.1600-0501.2002.130108.x.

Abstract

Resonance frequency analysis (RFA) has been used by several investigators to assess the boundary conditions of dental implants. The goal of the current study was to determine the vibrating behavior of a dental implant under various surrounding bone conditions. A 3D finite element (FE) model of a cylinder-type titanium implant was developed. In this model, the implant was embedded into a cubic section of bone. The model was first validated using a series of modal testing experiments. The effects of bony conditions on the resonance frequencies of the implant were computed with different bone types and bone densities. Our results show that the resonance frequency of the implant with type III surrounding bone decreased linearly (r = -0.996, P < 0.01) from 17.9 kHz (without loss in bone density) to 0.6 kHz (90% loss in bone density) when the bone densities were decreased. On the other hand, without bone loss, the highest resonance frequency value (36.1 kHz) was found when the implant was placed into type I surrounding bone. In contrast, the resonance frequency of the implant with type IV bone quality was found to be 9.9 kHz, which is almost four-fold less than that found in the type I model. These results suggest that RFA could serve as a non-invasive diagnostic tool for detecting the stability of dental implants during the healing stages and in subsequent routine follow-up care after treatment.

摘要

共振频率分析(RFA)已被多位研究人员用于评估牙种植体的边界条件。本研究的目的是确定牙种植体在各种周围骨条件下的振动行为。建立了圆柱型钛种植体的三维有限元(FE)模型。在该模型中,种植体被植入到一个立方体骨段中。该模型首先通过一系列模态测试实验进行了验证。计算了不同骨类型和骨密度对种植体共振频率的影响。我们的结果表明,当骨密度降低时,III型周围骨的种植体共振频率从17.9kHz(骨密度无损失)线性下降(r = -0.996,P < 0.01)至0.6kHz(骨密度损失90%)。另一方面,在无骨质流失的情况下,当种植体植入I型周围骨时,发现最高共振频率值为36.1kHz。相比之下,IV型骨质量的种植体共振频率为9.9kHz,几乎比I型模型中的共振频率低四倍。这些结果表明,RFA可作为一种非侵入性诊断工具,用于检测牙种植体在愈合阶段以及治疗后后续常规随访中的稳定性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验