Suppr超能文献

[Mechanical testing of implant properties of thoracoscopic implantation of ventral spinal stabilizing systems. Comparative study with the ISO/DIS 12189-2 corpectomy model and an improved synthetic model].

作者信息

Grupp T M, Beisse R, Potulski M, Marnay T, Beger J, Blömer W

机构信息

Aesculap AG & Co. KG, Forschung und Entwicklung, Am-Aesculap-Platz, 78532 Tuttlingen.

出版信息

Orthopade. 2002 Apr;31(4):406-12. doi: 10.1007/s00132-001-0284-7.

Abstract

A new modular anterior fixation system MACS TL (modular anterior construct system for the thoracic and lumbar spine) has been developed for use in thoracoscopic spondylodesis. This system demonstrates high angular stability and meets the surgical requirements for an endoscopic approach. The objective of the current study was fatigue testing of the MACS TL implant system using a corpectomy model according to ISO/DIS 12189-2 and a synthetic model recently developed by Kotani et al. [6]. The MACS TL system demonstrated good mechanical properties with a high stiffness compared to the published data reviewed. The importance of dynamic testing in a corpectomy model has been demonstrated by comparing the MACS TL plate system with an early prototype, which has not yet been clinically evaluated. The corpectomy model according to Kotani et al. offers an interesting alternative to the ISO/DIS 12189-2 test method for asymmetrically designed and antero-laterally positioned spinal implants due to the unconstrained ball joint.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验