Sato Eisuke, Nakashima Tsutomu, Lilly David J, Fausti Stephen A, Ueda Hiromi, Misawa Hayato, Uchida Yasue, Furuhashi Atsushi, Asahi Kiyomitsu, Naganawa Shinji
Department of Otorhinolaryngology, Nagoya University School of Medicine, Japan.
Laryngoscope. 2002 Sep;112(9):1642-6. doi: 10.1097/00005537-200209000-00021.
The purpose of this study was to study systematically some relationships between the resonance frequency of the middle-ear transmission system and the volume of the endolymphatic duct and sac in patients with an enlarged vestibular aqueduct (EVA).
Prospective study.
Thirteen patients (24 ears) with EVA, 17 subjects (29 ears) with normal hearing, and 17 patients (21 ears) with sensorineural hearing loss without EVA served as experimental subjects. Standard pure-tone audiometry, standard clinical tympanometry (using a 226-Hz probe tone), and multifrequency tympanometry were performed on each ear. Magnetic resonance imaging was used to determine the area of the cochlear modiolus and the volume of the endolymphatic duct and sac.
The audiometric configurations for most patients sloped downward from the low to the high frequencies. A significant air-bone gap was computed at each of these test frequencies. Multifrequency tympanometry yielded resonance frequencies for the patients with EVA that was significantly lower than those measured for the control subjects. In general, for patients with EVA, the resonance frequency of the middle ear system decreased as the volume of the endolymphatic duct and sac increased. This inverse relation was significant (correlation coefficient = -0.483, P =.0157). However, there was no correlation between resonance frequency and the degree of cochlea modiolar deficiency.
Clinically, our findings suggest that EVA probably should be included in the differential diagnosis for a patient who presents with a moderate to severe mixed hearing loss, a normal tympanogram at 226 Hz, and a resonance frequency that is abnormally low.
本研究旨在系统探究大前庭导水管(EVA)患者中耳传导系统的共振频率与内淋巴管及内淋巴囊体积之间的一些关系。
前瞻性研究。
选取13例(24耳)EVA患者、17例听力正常受试者(29耳)以及17例无EVA的感音神经性听力损失患者(21耳)作为实验对象。对每只耳朵进行标准纯音听力测定、标准临床鼓室导抗测试(使用226Hz探测音)以及多频鼓室导抗测试。采用磁共振成像确定蜗轴面积以及内淋巴管和内淋巴囊的体积。
大多数患者的听力图呈现从低频到高频向下倾斜的趋势。在每个测试频率下均计算出显著的气骨导差。多频鼓室导抗测试得出EVA患者的共振频率显著低于对照组受试者。总体而言,对于EVA患者,中耳系统的共振频率随着内淋巴管和内淋巴囊体积的增加而降低。这种负相关关系显著(相关系数 = -0.483,P = 0.0157)。然而,共振频率与蜗轴缺损程度之间无相关性。
在临床上,我们的研究结果表明,对于出现中度至重度混合性听力损失、226Hz鼓室导抗图正常且共振频率异常低的患者,鉴别诊断中可能应纳入EVA。