Real-time measurement and audit of radiation dose to patients undergoing computed radiography.

作者信息

Vano Eliseo, Fernandez Jose Miguel, Ten Jose Ignacio, Guibelalde Eduardo, Gonzalez Luciano, Pedrosa Cesar S A

机构信息

Department of Medical Physics, San Carlos University Hospital, Madrid, Spain.

出版信息

Radiology. 2002 Oct;225(1):283-8. doi: 10.1148/radiol.2243011331.

Abstract

A real-time patient dose monitoring system for auditing computed radiography is described. Technical data from each exposure and for every examination type are collected and sent by a network to a workstation, which calculates the moving average values of entrance skin dose and dose-area product from the 10 most recently examined patients. Comparison of averages with reference values generates warning messages if reference values are exceeded, prompting corrective action if necessary.