传统剂量与低剂量非增强头部CT图像质量的比较。

Comparison of image quality between conventional and low-dose nonenhanced head CT.

作者信息

Mullins Mark E, Lev Michael H, Bove Peter, O'Reilly Cara E, Saini Sanjay, Rhea James T, Thrall James H, Hunter George J, Hamberg Leena M, Gonzalez R Gilberto

机构信息

Neuroradiology Division, Massachusetts General Hospital and the Harvard Medical School, Boston, MA, USA.

出版信息

AJNR Am J Neuroradiol. 2004 Apr;25(4):533-8.

Abstract

BACKGROUND AND PURPOSE

Increasing use of CT for evaluating neurologic disease may expose patients to considerable levels of ionizing radiation. We compared the image quality of low-mAs head CT scans with that of conventional nonenhanced scans.

METHODS

Conventional head CT scans were obtained in 20 patients (all >65 years with history of non-CNS malignancy) by using a multidetector technique: 170 mA and 1-second scanning time (ie, 170 mAs), 140 kVp, table speed of 7.5 mm per rotation, pitch of 0.75, section thickness of 5 mm, and field of view of 25 mm. A limited volume helical data acquisition covering four 5-mm-thick images was obtained by using 90 mAs but otherwise the same parameters. Three neuroradiologists visually rated the resulting images for quality in a blinded comparison. Representative 1- to 4-mm(2) regions of interest were chosen in gray matter and white matter locations. Conspicuity and the contrast-to-noise ratio were analyzed. Statistical comparisons were done by using the Student t test.

RESULTS

Mean gray matter conspicuity was not significantly different between the 170- and 90-mAs groups (0.39 +/- 0.19 vs 0.41 +/- 0.03, P =.32). Mean gray matter contrast-to-noise ratio was approximately 22% higher with 170 mAs than with 90 mAs (1.77 +/- 0.52 vs 1.39 +/- 0.38, P =.005). All 90-mAs images were rated as having slightly greater image noise than the 170-mAs scans but with sufficient perceived resolution.

CONCLUSION

Although 90-mAs head CT images were moderately noisier than 170-mAs images, they were rated as having acceptable diagnostic quality.

摘要

背景与目的

越来越多地使用CT来评估神经系统疾病可能会使患者暴露于相当水平的电离辐射中。我们比较了低毫安秒头部CT扫描与传统非增强扫描的图像质量。

方法

对20例患者(均为年龄大于65岁且有非中枢神经系统恶性肿瘤病史者)采用多探测器技术进行传统头部CT扫描:170毫安、1秒扫描时间(即170毫安秒)、140千伏峰值、每旋转7.5毫米的床速、0.75的螺距、5毫米的层厚以及25毫米的视野。通过使用90毫安秒获取覆盖四张5毫米厚图像的有限容积螺旋数据采集,其他参数相同。三位神经放射科医生在双盲比较中对所得图像的质量进行视觉评分。在灰质和白质位置选择代表性的1至4平方毫米的感兴趣区域。分析了清晰度和对比噪声比。采用Student t检验进行统计学比较。

结果

170毫安秒组与90毫安秒组之间的平均灰质清晰度无显著差异(0.39±0.19对0.41±0.03,P = 0.32)。170毫安秒时的平均灰质对比噪声比比90毫安秒时约高22%(1.77±0.52对1.39±0.38,P = 0.005)。所有90毫安秒的图像被评为比170毫安秒的扫描图像噪声略大,但具有足够的可感知分辨率。

结论

尽管90毫安秒的头部CT图像比170毫安秒的图像噪声稍大,但它们被评为具有可接受的诊断质量。

相似文献

引用本文的文献

2
Comparing image quality of single- and dual-energy computed tomography of the brain.比较脑单能和双能 CT 成像的质量。
Neuroradiol J. 2020 Jun;33(3):259-266. doi: 10.1177/1971400920920785. Epub 2020 Apr 21.
8
Reduced-dose CT protocol for the assessment of cerebral vasospasm.用于评估脑血管痉挛的低剂量CT方案
Neuroradiology. 2015 Dec;57(12):1211-8. doi: 10.1007/s00234-015-1585-6. Epub 2015 Aug 28.

本文引用的文献

1
Multi-detector row CT: radiation dose characteristics.多排探测器CT:辐射剂量特征
Radiology. 2003 Mar;226(3):762-72. doi: 10.1148/radiol.2263020205.
4
Low-dose CT: how are we doing?低剂量CT:我们做得如何?
AJR Am J Roentgenol. 2003 Feb;180(2):303. doi: 10.2214/ajr.180.2.1800303.
7
Real-time measurement and audit of radiation dose to patients undergoing computed radiography.
Radiology. 2002 Oct;225(1):283-8. doi: 10.1148/radiol.2243011331.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验