Werner-Tutschku W, Lajtai G, Schmiedhuber G, Lang T, Pirkl C, Orthner E
Unfallchirurgische Abteilung, AKH-Wels, Germany.
Unfallchirurg. 2002 Oct;105(10):881-5. doi: 10.1007/s00113-002-0416-5.
In the period from 2/98 up to 6/99, fractures of the proximal femur of 70 patients (phi 79.2 a), were treated with a proximal femur nail (PFN((R))). The aim of this retrospective analysis was to evaluate intra- and perioperative complications. The reports of the operation and the anaesthesia documentation sheets have been worked out, all X-rays have been measured, and evaluated along exactly defined parameters. Problems or complications were found in 18 cases (25,7%). In 7 cases (10,0%) these problems could be solved during operation. The "Z-effect" seen in 5 cases (7,1%) and the "cut-out" of the sliding-hip-screw in 6 cases (8,6%) were the most frequent complications. The main reason was a bad primary reposition in varus with a CCD-Angle less than 125 degrees.
在1998年2月至1999年6月期间,70例(平均79.2岁)股骨近端骨折患者接受了股骨近端髓内钉(PFN®)治疗。本回顾性分析的目的是评估术中和围手术期并发症。整理了手术报告和麻醉记录单,对所有X线片进行了测量,并按照精确界定的参数进行了评估。18例(25.7%)发现了问题或并发症。7例(10.0%)的这些问题在手术中得到了解决。5例(7.1%)出现的“Z效应”和6例(8.6%)的滑动髋螺钉“穿出”是最常见的并发症。主要原因是初始内翻复位不佳,颈干角小于125度。